Handheld Echocardiography in a Clinical Practice Scenario: Concordances Compared to Standard Echocardiographic Reports.

作者信息

Blume Gustavo Gavazzoni, Lechinewski Luka David, Vieira Isabela Pedroza, Clausell Nadine, Bertinato Giovana Paludo, Machado-Júnior Paulo André Bispo, Berro Pedro Goulart, Moura Lidia Ana Zytynski, Tsang Teresa

机构信息

Division of Cardiovascular Diseases, Pontificial Catholic University of Paraná, Curitiba, Brazil.

Graphic Methods Sector, Holy House of Curitiba, Curitiba, Brazil.

出版信息

J Cardiovasc Imaging. 2022 Jan;30(1):25-34. doi: 10.4250/jcvi.2020.0241.

Abstract

BACKGROUND

The purpose of this study was to assess the utility of a handheld device (HH) used during common daily practice and its agreement with the results of a standard echocardiography study (STD) performed by experienced sonographers and echocardiographer.

METHODS

A prospective follow-up was conducted in an adult outpatient echocardiography clinic. Experienced sonographers performed the STD and an experienced echocardiographer performed the HH. STD included 2-dimensional images, Doppler and hemodynamics analysis. Hemodynamic assessment was not performed with the HH device because the HH does not include such technology. The images were interpreted by blinded echocardiographers, and the agreement between the reports was analyzed.

RESULTS

A total of 108 patients were included; and the concordance for left ventricle (LV) ejection fraction (EF), wall motion score index, LV and right ventricle (RV) function, RV size, and mitral and aortic stenosis was excellent with κ values greater than 0.80. Wall motion abnormalities had good concordance (κ value 0.78). The agreement for LV hypertrophy, mitral and aortic regurgitation was moderate, and tricuspid and pulmonary regurgitation agreements were low (κ values of 0.26 and 0.25, respectively).

CONCLUSIONS

In a daily practice scenario with experienced hands, HH demonstrated good correlation for most echocardiography indications, such as ventricular size and function assessment and stenosis valve lesion analyses.

摘要

背景

本研究的目的是评估日常临床实践中使用的手持设备(HH)的效用,以及它与经验丰富的超声检查技师和超声心动图医师进行的标准超声心动图检查(STD)结果的一致性。

方法

在一家成人门诊超声心动图诊所进行了一项前瞻性随访研究。经验丰富的超声检查技师进行STD,经验丰富的超声心动图医师进行HH检查。STD包括二维图像、多普勒和血流动力学分析。由于HH设备不具备此项技术,因此未使用HH设备进行血流动力学评估。图像由不知情的超声心动图医师解读,并分析报告之间的一致性。

结果

共纳入108例患者;左心室(LV)射血分数(EF)、室壁运动评分指数、LV和右心室(RV)功能、RV大小以及二尖瓣和主动脉瓣狭窄的一致性极佳,κ值大于0.80。室壁运动异常具有良好的一致性(κ值为0.78)。LV肥厚、二尖瓣和主动脉瓣反流的一致性为中等,三尖瓣和肺动脉瓣反流的一致性较低(κ值分别为0.26和0.25)。

结论

在由经验丰富的人员进行日常临床实践的情况下,HH在大多数超声心动图检查指征方面表现出良好的相关性,如心室大小和功能评估以及瓣膜狭窄病变分析。