The DMI PET/CT is a modular silicon photomultiplier-based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade.
PET performance was evaluated using the NEMA NU 2-2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality.
The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3-8.5% for 5 rings vs 1.9-6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions.
The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels.
DMI PET/CT 是一款基于模块化硅光电倍增管的扫描仪,其轴向视野(FOV)取决于环配置(3、4 或 5 个环),范围在 15 至 25cm 之间。该系统的新一代产品包括一个经过重新设计的探测器模块,具有改进的电子设备和额外的第 6 个环,将轴向 FOV 扩展到 30cm。我们报告了 6 环升级的第二代(Gen2)系统的性能评估,同时还报告了同一系统在升级前的 5 环配置的性能。
使用 NEMA NU 2-2018 标准评估 PET 性能,包括空间分辨率、灵敏度、图像质量、计数率性能、定时分辨率和图像配准精度。使用患者图像评估图像质量。
系统平均灵敏度测量值为 32.76 cps/kBq(与 5 环相比增加了约 47%,为 22.29 cps/kBq),噪声等效计数率峰值为 434.3 kcps,对应 23.6 kBq/mL(与第一代(Gen1)相比增加了约 60%,与 Gen2 5 环相比增加了 39%)。对比恢复率在 54.5%至 85.8%之间,与 5 环相似,而 6 环提供了更低的背景变异性(5 环为 2.3-8.5%,6 环为 1.9-6.8%)和更低的肺部误差(5 环为 4.0%,6 环为 3.16%)。在 1cm(3.79/4.26mm)和 10cm(4.29/4.55mm)处的横向/轴向半高全宽(FWHM)、散射分数(40.2%)、能量分辨率(9.63%)和时间飞行(TOF)分辨率(0 kBq/mL 时为 389.6 ps)与不同系统配置测量的先前报告值一致。与 5 环相比,6 环可获得更高的患者图像质量,并且即使在降低扫描时间的情况下也能保持图像质量,从而实现 WB 动态采集。
与 5 环配置相比,6 环 DMI 的更高灵敏度可能会提高临床图像的图像质量,同时减少扫描时间。此外,它还可以同样用于提高动态采集的时间采样和/或减少总扫描时间。相反,可以根据应用程序交换时间采样和扫描时间,以进一步降低较低水平的注射剂量。