BACKGROUND

The purpose of this study was to develop a swine model of surgically induced blood loss to evaluate the performances of a new autotransfusion system allowing red blood cells and platelets preservation while collecting, washing and concentrating hemorrhagic blood intraoperatively.

METHODS

Two types of surgically induced blood loss were used in 12 minipigs to assess system performance and potential animal complications following autotransfusion: a cardiac model (cardiopulmonary bypass) and a visceral model (induced splenic bleeding). Animal clinical and hematological parameters were evaluated at different time-points from before bleeding to the end of a 72-hour post-transfusion period and followed by a post-mortem examination. System performances were evaluated by qualitative and quantitative parameters.

RESULTS

All animals that received the autotransfusion survived. Minimal variations were seen on the red blood cell count, hemoglobin, hematocrit at the different sampling times. Coagulation tests failed to show any hypo or hypercoagulable state. Gross and histologic examination didn't reveal any thrombotic lesions. Performance parameters exceeded set objectives in both models: heparin clearance (≥ 90%), final heparin concentration (≤ 0.5 IU/mL), free hemoglobin washout (≥ 90%) and hematocrit (between 45% and 65%). The device treatment rate of diluted blood was over 80 mL/min.

CONCLUSIONS

In the present study, both animal models succeeded in reproducing clinical conditions of perioperative cardiac and non-cardiac blood loss. Sufficient blood was collected to allow evaluation of autotransfusion effects on animals and to demonstrate the system performance by evaluating its capacity to collect, wash and concentrate red blood cells and platelets. Reinfusion of the treated blood, containing not only concentrated red blood cells but also platelets, did not lead to any postoperative adverse nor thrombogenic events. Clinical and comparative studies need to be conducted to confirm the clinical benefit of platelet reinfusion.

摘要

背景

本研究旨在建立一种通过手术失血来评估新型血液回收系统性能的猪模型,该系统允许在手术过程中收集、洗涤和浓缩失血性血液的同时保存红细胞和血小板。

方法

使用两种类型的手术失血模型来评估 12 头小型猪的系统性能和潜在动物并发症,这些模型分别为心脏模型(心肺旁路)和内脏模型(诱导脾出血)。在从失血前到输血后 72 小时结束的不同时间点评估动物的临床和血液学参数,并进行死后检查。系统性能通过定性和定量参数进行评估。

结果

所有接受输血的动物均存活。在不同的采样时间,红细胞计数、血红蛋白和血细胞比容均有轻微变化。凝血试验未显示出任何低凝或高凝状态。大体和组织学检查未发现任何血栓病变。在两种模型中,性能参数均超过设定目标:肝素清除率(≥90%)、最终肝素浓度(≤0.5IU/mL)、游离血红蛋白洗脱率(≥90%)和血细胞比容(45%至 65%之间)。该设备处理稀释血液的速度超过 80 毫升/分钟。

结论

在本研究中,两种动物模型均成功地复制了围手术期心脏和非心脏失血的临床情况。收集到的血液足以评估输血对动物的影响,并通过评估其收集、洗涤和浓缩红细胞和血小板的能力来证明系统性能。输注处理后的血液,其中不仅含有浓缩的红细胞,还含有血小板,并未导致任何术后不良或血栓形成事件。需要进行临床和比较研究来确认血小板再输注的临床益处。