Acta Orthop. 2022 Apr 28;93:451-458. doi: 10.2340/17453674.2022.2267.
Ceramic liners may reduce early stability of uncemented acetabular components due to higher stiffness. However, the bone ingrowth capacities of porous trabecular titanium might compensate for this effect. This prospective randomized trial quantifies migration patterns of the Delta-TT cup, and compares polyethylene and ceramic liners.
Patients undergoing primary uncemented total hip arthroplasty with the Delta-TT cup and femoral stem with ceramic head were randomized to a polyethylene (n = 25) or ceramic (n = 28) liner. Radiostereometric analysis (RSA) radiographs, patient-reported hip function (HOOS-PS, OHS), and quality of life (EQ5D) were collected at baseline and 1.5, 3, 6, 12, and 24 months postoperatively. Model-based RSA was used to calculate 3D cup translation and rotation, and mixed models were used to compare effects over time between groups.
At 2 years follow-up, Delta-TT cups showed similar mean proximal translation of 0.56 mm (95% CI 0.38-0.75) in the ceramic (CE) group and 0.54 mm (0.30-0.77) in the polyethylene (PE) group, with a between group effect of 0.02 mm (-0.20-0.23). Most cup migration occurred in the first 1.5 to 3 months, stabilizing within 6 months. Any between-group effects were ≤ 0.30 mm for translation and ≤ 0.45° for rotation. Improvements in patient-reported hip function and quality of life were similar in both groups.
Regardless of liner type, Delta-TT cups showed some initial migration and stabilized within 6 months, which seems promising for long-term fixation in both cup-liner constructs.
陶瓷衬垫可能会由于更高的刚度而降低非骨水泥髋臼部件的早期稳定性。然而,多孔小梁钛的骨长入能力可能会弥补这种影响。本前瞻性随机试验定量测量了 Delta-TT 杯的迁移模式,并比较了聚乙烯和陶瓷衬垫。
接受 Delta-TT 杯和陶瓷头股骨柄的初次非骨水泥全髋关节置换术的患者,随机分为聚乙烯(n = 25)或陶瓷(n = 28)衬垫组。基线时以及术后 1.5、3、6、12 和 24 个月时,收集放射学体层摄影测量分析(RSA)射线照片、患者报告的髋关节功能(HOOS-PS、OHS)和生活质量(EQ5D)。使用基于模型的 RSA 计算 3D 杯的平移和旋转,并使用混合模型比较组间随时间的影响。
在 2 年的随访中,Delta-TT 杯在陶瓷(CE)组中的平均近端平移量为 0.56mm(95%CI 0.38-0.75),在聚乙烯(PE)组中的平均近端平移量为 0.54mm(0.30-0.77),组间效应为 0.02mm(-0.20-0.23)。大多数杯迁移发生在最初的 1.5 至 3 个月内,在 6 个月内稳定。任何组间效应的平移量均≤0.30mm,旋转量均≤0.45°。两组患者报告的髋关节功能和生活质量均有相似的改善。
无论衬垫类型如何,Delta-TT 杯都表现出一些初始迁移,并在 6 个月内稳定下来,这对于两种杯衬结构的长期固定似乎很有希望。