Glaucoma Investigation and Research Unit, the Royal Victorian Eye and Ear Hospital, Melbourne, VIC, Australia.
Department of Surgery Ophthalmology, University of Melbourne, Melbourne, VIC, Australia.
Eye (Lond). 2023 Apr;37(6):1184-1190. doi: 10.1038/s41433-022-02085-4. Epub 2022 May 16.
To establish a normative database and optimise parameters for personal-computer based perimetry via a web-application using circular contrast targets.
Online 24-degree 52-loci perimetry was delivered through a web-application using circular flickering contrast targets. Embedding contrast differentials within targets allows calculation of relative decibel (rdB) per 256-bit greyscale level differential. Target light-band maximum brightness colour was fixed, while the dark-band varied to achieve the desired rdB level. A staircase system was used with two reversals ranging from 0 to 36 rdB levels. Blind spot localisation at the start of the test was used to optimise viewing distance and subsequently count fixation losses. Gaze was maintained on a spinning golden star which moves mid-test to maximise sampling area. Patients performed the test to each eye separately using a computer in a darkened room at 40-45 cm.
158 eyes of 101 patients completed the Online Circular Contrast Perimetry (OCCP) test. Mean age was 62.9 ± 14.3 years old. Mean sensitivity reduced with age, at 1.0 relative decibel per decade. Mean sensitivity per locus correlated with standard automated perimetry (SAP) in a physiological hill of vision, with an average difference of 4.02 decibels (95% confidence interval (CI) = 3.77-4.27, p < 0.001) and good agreement between tests.
Online circular contract perimetry provides accurate perimetric testing with comparable results to standard automated perimetry.
通过使用网页应用程序中的圆形对比目标,为基于个人计算机的视野计建立规范数据库并优化参数。
通过使用网页应用程序中的圆形闪烁对比目标进行在线 24 度 52 个位置视野计检查。在目标中嵌入对比度差异可以计算每个 256 位灰度级差异的相对分贝 (rdB)。目标光带的最大亮度颜色固定,而暗带则变化以达到所需的 rdB 水平。使用 staircase 系统,从 0 到 36 rdB 水平进行两次反转。在测试开始时进行盲点定位,以优化观察距离,随后计算注视损失。测试过程中保持注视一个旋转的金色星星,该星星在中途移动以最大化采样区域。患者在黑暗的房间里,坐在距离电脑 40-45 厘米处,分别对每只眼睛进行测试。
101 名患者的 158 只眼完成了在线圆形对比视野计 (OCCP) 测试。平均年龄为 62.9 ± 14.3 岁。平均敏感度随年龄增长而降低,每 10 年降低 1 个相对分贝。每个位置的平均敏感度与标准自动视野计 (SAP) 在生理视野丘中相关,平均差异为 4.02 分贝(95%置信区间 (CI) = 3.77-4.27,p<0.001),两种测试之间具有良好的一致性。
在线圆形对比视野计提供了准确的视野计测试,结果与标准自动视野计相当。