Department of Physical Medicine and Rehabilitation, Centro Hospitalar Universitário São João, Porto, Portugal.
CINTESIS - Center for Health Technologies and Services Research, University of Porto, Porto, Portugal.
PLoS One. 2022 May 17;17(5):e0268254. doi: 10.1371/journal.pone.0268254. eCollection 2022.
The Handheld Dynamometer (HHD) has the potential to overcome some of the logistic and economic limitations of isokinetic dynamometers for measuring knee extension muscle strength. However, its reliability has not been fully assessed. The purpose of this study is to measure intra and inter-rater reliability of HHD for knee extension strength in patients receiving rehabilitation treatment, as well as to understand in which conditions is the reliability higher.
Twenty-nine patients admitted in an inpatient Physical Medicine and Rehabilitation unit were consecutively included in this cross-sectional study. Two experienced and two inexperienced physicians made two assessments of knee extension strength with HHD, separated by three hours. Intraclass Correlation Coefficients (ICC), absolute differences between assessments, and correlations between strength and functional variables were calculated.
Intra and inter-rater ICC were overall high (≥ 0.950 and 0.927, respectively). Higher values were found when average of two measurements were made for estimating intra-rater ICC (ICC = 0.978; 95%CI = 0.969-0.985) but not for inter-rater ICC. ICC were not statistically significantly different when calculated based on measurements performed by inexperienced physicians and experienced ones. There was a moderate correlation between strength and functional variables.
Handheld Dynamometer seems to be a reliable option to measure knee extension muscle strength, particularly when two measurements are performed and their average is reported.
手持测力计(HHD)有可能克服等速测力计在测量膝关节伸展肌力方面的一些物流和经济限制。然而,其可靠性尚未得到充分评估。本研究的目的是测量接受康复治疗的患者使用 HHD 测量膝关节伸展力量的内部和内部可靠性,以及了解在哪些情况下可靠性更高。
本横断面研究连续纳入 29 名在住院物理医学和康复科就诊的患者。两名经验丰富的医生和两名经验不足的医生使用 HHD 进行了两次膝关节伸展力量评估,间隔三小时。计算了组内相关系数(ICC)、两次评估之间的绝对差异以及力量与功能变量之间的相关性。
组内和组间 ICC 总体较高(分别为≥0.950 和 0.927)。当对两次测量的平均值进行组内 ICC 估计时,发现了更高的值(ICC=0.978;95%CI=0.969-0.985),但组间 ICC 并非如此。当根据经验不足的医生和经验丰富的医生进行的测量计算 ICC 时,ICC 没有统计学上的显著差异。力量与功能变量之间存在中度相关性。
手持测力计似乎是测量膝关节伸展肌力的可靠选择,特别是当进行两次测量并报告其平均值时。