IEEE Trans Biomed Circuits Syst. 2022 Aug;16(4):492-501. doi: 10.1109/TBCAS.2022.3181157. Epub 2022 Oct 12.
This paper presents an active impedance matching scheme that tries to optimize electrical power transfer and acoustic reflectivity in ultrasound transducers. Leveraging negative capacitance-based impedance matching would potentially improve the bandwidth and electrical power transfer while minimizing acoustic reflection of transducer elements and improve uniformity while reducing acoustic crosstalk of transducer arrays. A 16-element transceiver front-end is designed which employs an element-level active capacitive impedance cancellation scheme using an element-level negative impedance converter. The ASIC fabricated in 180-nm HVBCD technology provides high-voltage pulses up to 60 V consuming 3.6 mW and occupying 2.5 mm. The front-end ASIC is used with a 1-D capacitive micromachined ultrasonic transducer (CMUT) array and its acoustical reflectivity reduction and imaging capabilities have successfully been demonstrated through pulse-echo measurements and acoustic imaging experiments.
本文提出了一种主动阻抗匹配方案,旨在优化超声换能器中的电能传输和声反射。利用基于负电容的阻抗匹配有可能提高带宽和电能传输,同时最小化换能器元件的声反射,并提高换能器阵列的均匀性,同时降低声串扰。设计了一个 16 单元收发前端,它采用了一种基于元件级负阻抗转换器的元件级主动电容阻抗消除方案。该 ASIC 采用 180nm HVBCD 工艺制造,可提供高达 60V 的高压脉冲,功耗为 3.6mW,占用面积为 2.5mm。该前端 ASIC 与 1-D 电容式微机械超声换能器(CMUT)阵列一起使用,通过脉冲回波测量和声学成像实验成功地证明了其声反射降低和成像能力。