Shepherd A P, Riedel G L, Kiel J W, Haumschild D J, Maxwell L C
Am J Physiol. 1987 Jun;252(6 Pt 1):G832-9. doi: 10.1152/ajpgi.1987.252.6.G832.
Several laser-Doppler blood flowmeters are now commercially available; however, only one utilizes an infrared laser diode (Laserflo, TSI, St. Paul, MN). Because of this and other unique features such as its microprocessor-based signal analyzer, we evaluated this device's ability to measure tissue perfusion. Initially, we determined whether laser illumination directly affected the microvasculature. Intravital microscopic observations in the hamster cremaster muscle indicated that neither He-Ne nor infrared laser light affected the diameters of arterioles that were responsive to vasoactive agents. To test the flowmeter for linearity and repeatability, we used a rotating disk to simulate a light-scattering, flowing medium. The "flow" signal was highly correlated (r = 0.99) with the rotational velocity of the disk, was consistent among flow probes, and showed a high degree of reproducibility. The second model consisted of microsphere suspensions pumped through cuvettes. The laser-Doppler velocimeter (LDV) flow signal was linear with respect to pump output. With red blood cells in the perfusate, we examined the effects of blood oxygenation on the flowmeter's performance. The LDV flow signal was unaffected by changes in blood oxygenation. We evaluated linearity in vivo in isolated, perfused rat livers and in isolated canine gastric flaps. We observed linear relationships between total flow and laser-Doppler flow measured on the surface of the liver (r = 0.98) and in the gastric mucosa (r = 0.98), but the slopes of the relationships between total and local LDV flow showed considerable variability not noted in the in vitro studies.(ABSTRACT TRUNCATED AT 250 WORDS)
目前市场上有几种激光多普勒血流仪可供选择;然而,只有一种使用红外激光二极管(Laserflo,TSI,明尼苏达州圣保罗)。由于这一特点以及其他独特特性,如基于微处理器的信号分析仪,我们评估了该设备测量组织灌注的能力。最初,我们确定激光照射是否直接影响微脉管系统。在仓鼠提睾肌进行的活体显微镜观察表明,氦氖激光和红外激光均未影响对血管活性药物有反应的小动脉直径。为了测试血流仪的线性和可重复性,我们使用旋转盘模拟光散射、流动的介质。“血流”信号与盘的旋转速度高度相关(r = 0.99),在不同的血流探头之间具有一致性,并且显示出高度的可重复性。第二个模型由通过比色皿泵送的微球悬浮液组成。激光多普勒测速仪(LDV)的血流信号与泵输出呈线性关系。在灌注液中加入红细胞后,我们研究了血液氧合对血流仪性能的影响。LDV血流信号不受血液氧合变化的影响。我们在离体灌注的大鼠肝脏和离体犬胃瓣中评估了体内的线性关系。我们观察到肝脏表面测量的总血流与激光多普勒血流之间(r = 0.98)以及胃黏膜中(r = 0.98)存在线性关系,但总血流与局部LDV血流之间关系的斜率显示出相当大的变异性,这在体外研究中未观察到。(摘要截短于250字)