Evaluation of volumetric blood collection devices for the measurement of phenylalanine and tyrosine to monitor patients with phenylketonuria.

作者信息

Carling Rachel S, Emmett Erin C, Moat Stuart J

机构信息

GKT School Medical Education, Kings College London, UK; Biochemical Sciences, Viapath, Guys & St Thomas' NHSFT, London, UK.

Biochemical Sciences, Viapath, Guys & St Thomas' NHSFT, London, UK.

出版信息

Clin Chim Acta. 2022 Oct 1;535:157-166. doi: 10.1016/j.cca.2022.08.005. Epub 2022 Aug 20.

Abstract

BACKGROUND

Measurement of dried blood spot (DBS) phenylalanine (Phe) is central to the monitoring of patients with phenylketonuria. However, the volume and hematocrit (Hct) of the blood applied to conventional DBS cards significantly affects analytical results. Volumetric blood collection devices are reported to be more accurate, precise and less prone to Hct effects.

METHODS

Accuracy, imprecision, effect of blood volume and Hct were evaluated for measurement of Phe and tyrosine using three volumetric devices and compared with the conventional PerkinElmer-226 filter-paper collection devices. i.e. conventional DBS cards. Applicability for use in a clinical laboratory was assessed qualitatively.

RESULTS

Blood volume did not impact on the performance of the volumetric devices; however, significant biases were observed with the conventional DBS card. A higher Hct introduced unacceptable bias for Neoteryx-Mitra and conventional DBS card. All devices had a mean relative standard deviation (RSD) ≤ 4.1 %, except for the Neoteryx-Mitra (≤ 6.2 %). Relative to liquid blood, the mean biases of Phe for the various devices were -5.1 (HemaXis-DB10), -7.8 (Capitainer-qDBS), -12.0 (Neoteryx-Mitra) and -32.6 % (conventional DBS card).

CONCLUSIONS

Introducing volumetric collection devices will overcome the significant pre-analytical issues associated with conventional DBS collection and improve the biochemical monitoring of patients with PKU.

摘要

背景

干血斑(DBS)苯丙氨酸(Phe)检测是苯丙酮尿症患者监测的核心。然而,涂覆在传统DBS卡上的血液体积和血细胞比容(Hct)会显著影响分析结果。据报道,定量采血设备更准确、精确,且受Hct影响较小。

方法

使用三种定量设备评估Phe和酪氨酸检测的准确性、不精密度、血容量和Hct的影响,并与传统的珀金埃尔默-226滤纸采集设备(即传统DBS卡)进行比较。定性评估其在临床实验室的适用性。

结果

血容量对定量设备的性能没有影响;然而,传统DBS卡观察到显著偏差。较高的Hct对Neoteryx-Mitra和传统DBS卡引入了不可接受的偏差。除Neoteryx-Mitra(≤6.2%)外,所有设备的平均相对标准偏差(RSD)≤4.1%。相对于液态血,各种设备的Phe平均偏差分别为-5.1(HemaXis-DB10)、-7.8(Capitainer-qDBS)、-12.0(Neoteryx-Mitra)和-32.6%(传统DBS卡)。

结论

引入定量采集设备将克服与传统DBS采集相关的重大分析前问题,并改善苯丙酮尿症患者的生化监测。