Francis I. Proctor Foundation, University of California, San Francisco, CA, USA.
Narayana Nethralaya Eye Hospital, Bangalore, India.
Eye (Lond). 2023 Jun;37(8):1690-1695. doi: 10.1038/s41433-022-02223-y. Epub 2022 Sep 5.
BACKGROUND/OBJECTIVES: Tablet-based perimetry could be used to test for glaucomatous visual field defects in settings without easy access to perimeters, although few studies have assessed diagnostic accuracy of tablet-based tests. The goal of this study was to determine the diagnostic accuracy of iPad perimetry using the visualFields Easy application.
SUBJECTS/METHODS: This was a prospective, cross-sectional study of patients undergoing their first Humphrey Field Analyser (HFA) visual field test at a glaucoma clinic in India. Participants underwent 24-2 SITA Standard HFA testing and iPad-based perimetry with the visualFields Easy application. Reference standards for both visual field loss and suspected glaucoma were determined by ophthalmologist review of HFA results and optic disc photographs. Receiver operating characteristic curves were constructed to assess diagnostic accuracy at various test thresholds.
203 eyes from 115 participants were included, with 82 eyes classified as moderate or worse glaucoma. iPad perimetry had an area under the receiver operating characteristic (AUROC) curve of 0.64 (95% CI 0.57 to 0.71) for detection of any visual field defect relative to HFA and an AUROC of 0.68 (0.59 to 0.76) for detection of moderate or worse glaucoma relative to ophthalmologist examination. At a set specificity of 90%, the sensitivity of iPad perimetry for detection of moderate or worse glaucoma was 35% (22-48%).
iPad perimetry using the visualFields Easy application had inadequate diagnostic accuracy to be used as a screening tool for glaucoma in this South Indian population.
背景/目的:在难以获得视野计的环境中,可以使用基于平板电脑的视野计来测试青光眼的视野缺陷,尽管很少有研究评估基于平板电脑的测试的诊断准确性。本研究的目的是确定使用 visualFields Easy 应用程序的 iPad 视野计的诊断准确性。
受试者/方法:这是一项在印度一家青光眼诊所进行的前瞻性、横断面研究,纳入了首次接受 Humphrey 视野分析仪(HFA)视野测试的患者。参与者接受了 24-2 SITA 标准 HFA 测试和基于 visualFields Easy 应用程序的 iPad 视野计检查。HFA 结果和视盘照片由眼科医生审查来确定视野丧失和疑似青光眼的参考标准。构建了受试者工作特征曲线来评估不同测试阈值下的诊断准确性。
纳入了 115 名参与者的 203 只眼,其中 82 只眼被归类为中度或更严重的青光眼。与 HFA 相比,iPad 视野计检测任何视野缺陷的受试者工作特征(ROC)曲线下面积(AUROC)为 0.64(95%CI 0.57 至 0.71),检测中度或更严重青光眼的 AUROC 为 0.68(0.59 至 0.76)。在设定的特异性为 90%时,iPad 视野计检测中度或更严重青光眼的敏感性为 35%(22-48%)。
在该印度南部人群中,使用 visualFields Easy 应用程序的 iPad 视野计的诊断准确性不足以作为青光眼的筛查工具。