Pentacam 眼前节分析系统与超声角膜测厚仪测量中央角膜厚度的可重复性、重复性及相关性

Reproducibility, Repeatability, and Correlation of Central Corneal Thickness Measurement with the Pentacam Scheimpflug System and Ultrasound Pachymetry.

作者信息

Macias-Rodriguez Yolanda, Ramos-Dávila Eugenia M, Ruiz-Lozano Raul E, Reyes-Arena Jose V, Rivera-Alvarado I Jocelyn, Hernandez-Camarena Julio C, Rodriguez-Garcia Alejandro

机构信息

Instituto de Oftalmologia y Ciencias Visuales, Tecnologico de Monterrey, Escuela de Medicina y Ciencias de la Salud, Monterrey, Mexico.

出版信息

Klin Monbl Augenheilkd. 2024 Nov;241(11):1238-1244. doi: 10.1055/a-1938-4491. Epub 2022 Sep 7.

Abstract

BACKGROUND/AIM: The importance of an accurate determination of central corneal thickness (CCT) relies on its diagnostic and therapeutic implications in glaucoma, corneal ectasias, corneal edema and endothelial function monitoring, and refractive surgery suitability, among others. We aimed to analyze the repeatability, reproducibility, correlation, and laterality variations of CCT measurements performed with the Pentacam HR and the standard of care ultrasound A-scan (USP).

METHODS

A cross-sectional study including CCT measurements of healthy individuals was performed by three independent examiners with the Pentacam HR and USP. Intra-observer and inter-observer variations were calculated with intra-class correlation coefficients (ICCs). Bland-Altman plots and 95% limits of agreement (95% LoA) were used to assess the agreement between devices. Linear correlation was calculated with Pearson's coefficient.

RESULTS

Thirty individuals (60 eyes), including 10 (33.3%) men and 20 (66.6%) women, with a mean age of 30.0 ± 9.1 years, were studied. No statistical differences were found in CCT measurements between Pentacam HR (range 500 - 609 µm) and USP (range 498 - 628 µm). There was a high degree of correlation in repeatability and reproducibility of each independent device (ICC > 0.90). Pearson's correlation between 1 vs. 2, 2 vs. 3, and 3 vs. 1 Pentacam HR attempts were 0.914, 0.958, and 0.925, respectively (p < 0.001). Corresponding results for USP were 0.957, 0.957, and 0.943 (p < 0.001). The Pentacam HR tended to overestimate CCT by a mean difference of 3.77 µm (95% LoA, - 24.9 - 18.4). Right eyes were also overestimated (- 3.6 ± 14.1 µm) with the Pentacam HR device, whereas left eyes were underestimated (1.3 ± 11.1 µm).

CONCLUSIONS

The Pentacam HR device provides reliable operator-independent estimates of CCT. Right eyes exhibited a tendency to overestimate with the Pentacam HR. We suspect this difference is due to USP underestimation related to patients' position while performing the study. In clinically relevant scenarios, performing a third measurement and cautiously measuring right eyes can provide higher accuracy.

摘要

背景/目的:准确测定中央角膜厚度(CCT)的重要性在于其在青光眼、角膜扩张、角膜水肿、内皮功能监测以及屈光手术适用性等方面的诊断和治疗意义。我们旨在分析使用Pentacam HR和标准护理超声A扫描(USP)进行的CCT测量的可重复性、再现性、相关性和双眼差异。

方法

由三名独立检查者使用Pentacam HR和USP对健康个体进行CCT测量的横断面研究。使用组内相关系数(ICC)计算观察者内和观察者间的差异。使用Bland-Altman图和95%一致性界限(95% LoA)评估设备之间的一致性。使用Pearson系数计算线性相关性。

结果

研究了30名个体(60只眼),包括10名(33.3%)男性和20名(66.6%)女性,平均年龄为30.0±9.1岁。Pentacam HR(范围500 - 609 µm)和USP(范围498 - 628 µm)之间的CCT测量未发现统计学差异。每个独立设备的可重复性和再现性具有高度相关性(ICC > 0.90)。Pentacam HR的第1次与第2次、第2次与第3次、第3次与第1次测量之间的Pearson相关性分别为0.914、0.958和0.925(p < 0.001)。USP的相应结果为0.957、0.957和0.943(p < 0.001)。Pentacam HR倾向于高估CCT,平均差异为3.77 µm(95% LoA,-24.9 - 18.4)。使用Pentacam HR设备时右眼也被高估(-3.6±14.1 µm),而左眼被低估(1.3±11.1 µm)。

结论

Pentacam HR设备可提供可靠的、与操作者无关的CCT估计值。Pentacam HR对右眼有高估倾向。我们怀疑这种差异是由于在进行研究时与患者体位相关的USP低估所致。在临床相关情况下,进行第三次测量并谨慎测量右眼可提供更高的准确性。