Phan C Q, Tremper K K, Lee S E, Barker S J
J Clin Monit. 1987 Jul;3(3):149-54. doi: 10.1007/BF01695936.
This study compares two noninvasive techniques for monitoring the partial pressure of carbon dioxide (PCO2) in 24 anesthetized adult patients. End-tidal PCO2 (PetCO2) and transcutaneous PCO2 (PtcCO2) were simultaneously monitored and compared with arterial PCO2 (PaCO2) determined by intermittent analysis of arterial blood samples. PETCO2 and PtcCO2 values were compared with PaCO2 values corrected to patient body temperature (PaCO2T) and PaCO2 values determined at a temperature of 37 degrees C (PaCO2). Linear regression was performed along with calculations of the correlation coefficient (r), bias, and precision of the four paired variables: PETCO2 versus PaCO2 and PaCO2T (n = 211), and PtcCO2 versus PaCO2 and PaCO2T (n = 233). Bias is defined as the mean difference between paired values, whereas precision is the standard deviation of the difference. The following values were found for r, bias, and +/- precision, respectively. PETCO2 versus PaCO2: 0.67, -7.8 mm Hg, +/- 6.1 mm Hg; PETCO2 versus PaCO2T: 0.73, -5.8 mm Hg, +/- 5.9 mm Hg; PtcCO2 versus PaCO2: 0.87, -1.6 mm Hg, +/- 4.3 mm Hg; PtcCO2 versus PaCO2T: 0.84, +0.7 mm Hg, +/- 4.8 mm Hg. Although each of these PCO2 variables is physiologically different, there is a significant correlation (P less than 0.001) between the noninvasively monitored values and the blood gas values. Temperature correction of the arterial values (PaCO2T) slightly improved the correlation, with respect to PETCO2, but it had the opposite effect for PtcCO2. In this study, the chief distinction between these two noninvasive monitors was that PETCO2 had a large negative bias, whereas PtcCO2 had a small bias. We conclude from these data that PtcCO2 may be used to estimate PaCO2 with an accuracy similar to that of PETCO2 in anesthetized patients.
本研究比较了两种用于监测24例成年麻醉患者二氧化碳分压(PCO2)的非侵入性技术。同时监测呼气末二氧化碳分压(PetCO2)和经皮二氧化碳分压(PtcCO2),并与通过间歇性分析动脉血样本测定的动脉二氧化碳分压(PaCO2)进行比较。将PetCO2和PtcCO2值与校正至患者体温的PaCO2值(PaCO2T)以及在37摄氏度温度下测定的PaCO2值(PaCO2)进行比较。对四对配对变量进行线性回归分析,并计算相关系数(r)、偏差和精密度:PetCO2与PaCO2和PaCO2T(n = 211),以及PtcCO2与PaCO2和PaCO2T(n = 233)。偏差定义为配对值之间的平均差异,而精密度是差异的标准差。分别得到以下r、偏差和±精密度值。PetCO2与PaCO2:0.67,-7.8 mmHg,±6.1 mmHg;PetCO2与PaCO2T:0.73,-5.8 mmHg,±5.9 mmHg;PtcCO2与PaCO2:0.87,-1.6 mmHg,±4.3 mmHg;PtcCO2与PaCO2T:0.84,+0.7 mmHg,±4.8 mmHg。尽管这些PCO2变量在生理上各有不同,但非侵入性监测值与血气值之间存在显著相关性(P < 0.001)。动脉值(PaCO2T)的温度校正对于PetCO2而言,稍微改善了相关性,但对PtcCO2则产生了相反的效果。在本研究中,这两种非侵入性监测仪的主要区别在于,PetCO2存在较大的负偏差,而PtcCO2的偏差较小。从这些数据我们得出结论,在麻醉患者中,PtcCO2可用于估计PaCO2,其准确性与PetCO2相似。