The Berthold LB950 Automatic Luminescence Analyser was used to estimate bacterial adenosine triphosphate in urine. The system provided a rapid (15 min) and fully automated screening test for bacteriuria at the 10(5) CFU/ml level. Bioluminescence results for 1040 urines were compared with viable counts using two reference culture methods and frequency distributions of bacterial counts and adenosine triphosphate levels were calculated. With a specificity of 79% the automated method showed a sensitivity of 84% using a pour plate reference count and 91% using a standard loop reference count. When contaminated urines were excluded the sensitivity improved to 98%. The automated bioluminescence test, though expensive, was shown to work well with good quality specimens.

摘要

使用贝托尔德LB950自动发光分析仪来估计尿液中的细菌三磷酸腺苷。该系统针对每毫升10⁵ 菌落形成单位(CFU/ml)水平的菌尿症提供了一种快速(15分钟)且完全自动化的筛查测试。使用两种参考培养方法将1040份尿液的生物发光结果与活菌计数进行比较,并计算细菌计数和三磷酸腺苷水平的频率分布。该自动化方法的特异性为79%,使用倾注平板参考计数时灵敏度为84%,使用标准接种环参考计数时灵敏度为91%。当排除受污染的尿液时,灵敏度提高到98%。自动化生物发光测试虽然昂贵,但在处理高质量标本时效果良好。