Ayna Mustafa, Spille Johannes, Açil Yahya, Weitkamp Jan-Tobias, Wiltfang Jörg, Esen Cemal, Gülses Aydin
Department of Periodontology, Bonn University, 53111 Bonn, Germany.
Department of Oral and Maxillofacial Surgery, Christian Albrecht University, 24105 Kiel, Germany.
Materials (Basel). 2022 Nov 8;15(22):7889. doi: 10.3390/ma15227889.
The aim of the current experimental study was to comparatively assess the surface alterations in titanium and titanium-zirconium alloy implants in terms of thread pitch topography after irradiation with an Er:YAG laser, which is recommended in the literature for its sterilizing effect in the treatment of contaminated implant surfaces. Roxolid and SLA (Sand-blasted, Large-grit, Acid-etched) implants from Straumann company with the same macro topography were investigated. The surface treatment was carried out using a wavelength of 2940 nm, 60 s irradiation time, a frequency of 10 Hz, and energies between 120 mJ and 250 mJ. The alterations were quantitatively analyzed by conducting roughness analysis via white light interferometry and qualitatively using SEM images. Roxolid could particularly maintain its surface topography at a level of 160 mJ. At an energy level of 250 mJ, the surface properties of the pitch could be significantly altered for the first time. Compared to the Standard Plus dental implants studied, no distinct removal of the material from the surface was detected. The alloy properties of Roxolid confirm the manufacturer's statement in terms of stability and could offer advantages in peri-implantitis management if decontamination has been selected. However, as a part of a respective strategy, smoothening of a Roxolid implant surface requires a significantly higher energy level compared to SLA-Standard dental implants.
当前实验研究的目的是,在用铒钇铝石榴石激光照射后,根据螺距形貌对钛及钛锆合金种植体的表面变化进行比较评估,鉴于文献中推荐该激光对受污染种植体表面具有杀菌作用。研究了来自士卓曼公司具有相同宏观形貌的Roxolid和SLA(喷砂、大颗粒、酸蚀)种植体。表面处理采用波长2940 nm、照射时间60 s、频率10 Hz以及120 mJ至250 mJ之间的能量进行。通过白光干涉测量法进行粗糙度分析对变化进行定量分析,并使用扫描电子显微镜图像进行定性分析。Roxolid在160 mJ的能量水平下能够特别地保持其表面形貌。在250 mJ的能量水平下,螺距的表面特性首次可能发生显著变化。与所研究的标准加强型牙科种植体相比,未检测到表面有明显的材料去除。Roxolid的合金特性在稳定性方面证实了制造商的说法,并且如果选择进行去污处理,在种植体周围炎的管理方面可能具有优势。然而,作为相应策略的一部分,与SLA标准牙科种植体相比,Roxolid种植体表面的平滑处理需要显著更高的能量水平。