Sensors (Basel). 2022 Dec 29;23(1):377. doi: 10.3390/s23010377.
Consideration of pelvic mobility when positioning implants for total hip arthroplasty (THA) has been shown to reduce the risk of complications such as dislocation, squeaking and excessive wear. We aim to test the repeatability of pelvic tilt measurements taken between three positions (standing, flexed-seated and step-up) by an inertial measurement unit (IMU) and hence, evaluate their reliability in screening for high pelvic mobility in patients undergoing THA. The repeated IMU measurements of pelvic tilt were analysed for consistency and compared with measures taken by x-ray analysis. Our study showed greater variation in measures taken by the IMU particularly in the flexed-seated position. The patient's pelvic tilt in this position negatively correlated with their mid-back angle, suggesting the posture of the patient is a source of variation in the flexed-seated position if not kept consistent during assessments. IMUs were overall able to produce accurate and reliable measurements of pelvic tilt; however, protocols will need to be adjusted to factor in a patient's mid-back angle when taking future readings.
在进行全髋关节置换术 (THA) 时,考虑骨盆活动度已被证明可以降低脱位、异响和过度磨损等并发症的风险。我们旨在通过惯性测量单元 (IMU) 测试三种体位(站立位、屈髋坐位和台阶位)下骨盆倾斜度测量的可重复性,并评估其在筛查 THA 患者骨盆高活动性中的可靠性。对 IMU 测量的骨盆倾斜度进行了一致性分析,并与 X 射线分析测量值进行了比较。我们的研究表明,IMU 测量值的变化较大,尤其是在屈髋坐位时。该体位下患者的骨盆倾斜度与中背部角度呈负相关,这表明如果在评估过程中不保持一致,患者的姿势是该体位下产生变化的一个来源。IMU 总体上能够准确可靠地测量骨盆倾斜度;但是,未来在进行测量时,需要根据患者的中背部角度来调整方案。