Department of Diagnostic and Interventional Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Department of Diagnostic and Interventional Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Z Med Phys. 2024 Nov;34(4):596-609. doi: 10.1016/j.zemedi.2023.01.002. Epub 2023 Jan 28.
To test intra-arterial spin labeling (iASL) using active guiding catheters for myocardial perfusion measurements during magnetic resonance (MR)-guided interventions in a pig study.
In this work, a single-loop radiofrequency (RF) coil at the tip of a 6F active coronary catheter was used as a transmit coil for local spin labeling. The transmit magnetic RF field (B) of the coil and the labeling efficiency were determined, and iASL was tested in two pigs after the catheter was engaged in the aortic root, the ostium of the left coronary artery (LCA) under MR-guidance. The iASL effect was assessed by the signal difference between spin-labeling On and control (spin-labeling OFF) images, and in a cross-correlation between ON/Off states of spin-labeling a binary labeling paradigm. In addition, quantitative myocardial perfusion was calculated from the iASL experiments.
The maximum B in the vicinity of the catheter coil was 2.1 µT. A strong local labeling effect with a labeling efficiency of 0.45 was achieved with iASL both in vitro and in vivo. In both pigs, the proximal myocardial segments supplied by the LCA showed significant labelling effect up to distances of 60 mm from the aortic root with a relative signal difference of (3.14 ± 2.89)% in the first and (3.50 ± 1.25)% in the second animal. The mean correlation coefficients were R = 0.63 ± 0.22 and 0.42 ± 0.16, respectively. The corresponding computed myocardial perfusion values in this region of the myocardium were similar to those obtained with contrast perfusion methods ((1.2 ± 1.1) mL/min/g and (0.8 ± 0.6) mL/min/g).
The proposed iASL method demonstrates the feasibility of selective myocardial perfusion measurements during MR-guided coronary interventions, which with further technical improvements may provide an alternative to exogenous contrast-based perfusion. Due to the invasive nature of the iASL method, it can potentially be used in concert with MRI-guided coronary angioplasty.
在猪模型的磁共振(MR)引导介入中,使用主动引导导管进行心肌灌注测量,验证动脉自旋标记(iASL)技术的可行性。
在这项工作中,我们将单匝射频(RF)线圈置于 6F 主动冠状动脉导管的尖端,作为局部自旋标记的发射线圈。我们测定了线圈的发射磁场(B)和标记效率,并在导管在 MR 引导下进入主动脉根部和左冠状动脉(LCA)开口后,在两只猪身上进行了 iASL 测试。我们通过自旋标记 ON 与对照(自旋标记 OFF)图像之间的信号差异以及自旋标记 ON/OFF 状态之间的交叉相关来评估 iASL 效果,还通过 iASL 实验计算了定量心肌灌注。
导管线圈附近的最大 B 值为 2.1μT。在体外和体内,iASL 均实现了强烈的局部标记效果,标记效率为 0.45。在两只猪中,LCA 供血的近端心肌节段在距离主动脉根部 60mm 处显示出明显的标记效果,第一只动物的相对信号差异为(3.14±2.89)%,第二只动物为(3.50±1.25)%。平均相关系数分别为 R=0.63±0.22 和 0.42±0.16。相应的心肌灌注值与对比灌注方法相似((1.2±1.1)mL/min/g 和(0.8±0.6)mL/min/g)。
所提出的 iASL 方法证明了在 MR 引导的冠状动脉介入期间进行选择性心肌灌注测量的可行性,该方法进一步的技术改进可能为外源性对比灌注提供替代方法。由于 iASL 方法具有侵入性,因此它可能与 MRI 引导的冠状动脉成形术联合使用。