Aktiia SA, Neuchâtel, Switzerland.
Guangdong Transtek Medical Electronics, Zhongshan, China.
Blood Press Monit. 2023 Apr 1;28(2):109-112. doi: 10.1097/MBP.0000000000000639. Epub 2023 Feb 17.
Assess the accuracy and precision of the Aktiia initialization oscillometric upper-arm cuff device (Aktiia SA, Neuchâtel, Switzerland) for home blood pressure (BP) monitoring in the general population according to the American National Standards Institute / Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard.
Three trained observers validated BP measurements performed using the Aktiia cuff versus BP measurements performed using a standard mercury sphygmomanometer. Two ISO 81060-2 criteria were used to validate the Aktiia cuff. Criterion 1 evaluated, for both SBP and DBP, whether the mean error between BP readings performed by the Aktiia cuff and auscultation was ≤±5 mmHg, and whether the SD of the error was ≤8 mmHg. Criterion 2 assessed whether, for the SBP and DBP of each individual subject, the SD of the averaged paired determinations per subject of the Aktiia cuff and of the auscultation met the criteria listed in the table of Averaged Subject Data Acceptance.
Mean differences between the Aktiia cuff and the standard mercury sphygmomanometer (criterion 1) were 1.3 ± 7.11 mmHg for SBP and -0.2 ± 5.46 mmHg for DBP. The SD of the averaged paired differences per subject (criterion 2) was 6.55 mmHg for SBP and 5.15 mmHg for DBP.
Aktiia initialization cuff complies with the requirements of the ANSI/AAMI/ISO guidelines and can be safely recommended for BP measurements in the adult population.
根据美国国家标准协会/医学仪器促进协会/国际标准化组织(ANSI/AAMI/ISO)81060-2:2013 标准,评估 Aktiia 初始化测振法上臂袖带设备(Aktiia SA,纳沙泰尔,瑞士)在普通人群中进行家庭血压(BP)监测的准确性和精密度。
三名经过培训的观察者对使用 Aktiia 袖带进行的 BP 测量与使用标准水银血压计进行的 BP 测量进行了验证。使用了两个 ISO 81060-2 标准来验证 Aktiia 袖带。标准 1 评估了 Aktiia 袖带测量的收缩压和舒张压的平均误差是否在±5mmHg 以内,误差的标准差是否在 8mmHg 以内。标准 2 评估了每个个体的收缩压和舒张压,Aktiia 袖带和听诊法的每个个体的平均配对测量的标准差是否符合表中列出的平均个体数据接受标准。
Aktiia 袖带与标准水银血压计之间的平均差值(标准 1)为收缩压 1.3±7.11mmHg,舒张压-0.2±5.46mmHg。每个个体的平均配对差值的标准差(标准 2)为收缩压 6.55mmHg,舒张压 5.15mmHg。
Aktiia 初始化袖带符合 ANSI/AAMI/ISO 指南的要求,可安全推荐用于成人的血压测量。