The aim of this study was to evaluate the measurements of corneal higher-order aberrations (HOAs) obtained by a new anterior segment optical coherence tomography (OCT) technique combined with a Placido topographer (the MS-39 device) in eyes with prior small-incision lenticule extraction (SMILE) and compare them to the measurements obtained by a Scheimpflug camera combined with a Placido topographer (the Sirius device).
A total of 56 eyes (56 patients) were included in this prospective study. Corneal aberrations were analyzed for the anterior, posterior, and total cornea surfaces. Within-subject standard deviation (S), test-retest repeatability (TRT), and intraclass correlation coefficient (ICC) were used to assess the intraobserver repeatability and interobserver reproducibility. The differences were evaluated by paired t-test. Bland-Altman plots and 95% limits of agreement (95% LoA) were used to evaluate the agreement.
High repeatability was observed for anterior and total corneal parameters, with S value < 0.07, TRT ≤ 0.16, and ICCs > 0.893, but not trefoil. For the posterior corneal parameters, ICCs varied from 0.088 to 0.966. Regarding interobserver reproducibility, all S values were ≤ 0.04 and TRT ≤ 0.11. ICCs ranged from 0.846 to 0.989, from 0.432 to 0.972, and from 0.798 to 0.985 for the anterior, total, and posterior corneal aberrations parameters, respectively. The mean difference in all aberrations was ≤ 0.05 μm. All parameters showed a narrow 95% LoA.
The MS-39 device achieved high precision in both anterior and total corneal measurements; the precision of posterior corneal higher-order RMS, astigmatism II, coma, and trefoil was lower. The two technologies used by the MS-39 and Sirius devices can be used interchangeably for measuring corneal HOAs after SMILE.
本研究的目的是评估一种新的前段光学相干断层扫描(OCT)技术与普拉西多角膜地形图仪(MS-39设备)相结合,在既往接受过小切口透镜切除术(SMILE)的眼睛中测量角膜高阶像差(HOA)的情况,并将其与结合了普拉西多角膜地形图仪的Scheimpflug相机(Sirius设备)所获得的测量结果进行比较。
本前瞻性研究共纳入56只眼(56例患者)。对角膜前表面、后表面和整个角膜表面的像差进行分析。采用受试者内标准差(S)、重测重复性(TRT)和组内相关系数(ICC)来评估观察者内重复性和观察者间再现性。采用配对t检验评估差异。使用Bland-Altman图和95%一致性界限(95% LoA)来评估一致性。
角膜前表面和整个角膜参数具有高重复性,S值<0.07,TRT≤0.16,ICC>0.893,但三叶形像差除外。对于角膜后表面参数,ICC范围为0.088至0.966。关于观察者间再现性,所有S值均≤0.04,TRT≤0.11。角膜前表面、整个角膜和后表面像差参数的ICC分别为0.846至0.989、0.432至0.972和0.798至0.985。所有像差的平均差异≤0.05μm。所有参数均显示出较窄的95% LoA。
MS-39设备在角膜前表面和整个角膜测量中均实现了高精度;角膜后表面高阶均方根、散光II、彗差和三叶形像差的测量精度较低。MS-39和Sirius设备所使用的两种技术在SMILE术后测量角膜HOA时可互换使用。