Suppr超能文献

机械通气期间通过热稀释法改善心输出量估计值

Improvement of cardiac output estimation by the thermodilution method during mechanical ventilation.

作者信息

Jansen J R, Versprille A

出版信息

Intensive Care Med. 1986;12(2):71-9. doi: 10.1007/BF00254515.

Abstract

The reliability of cardiac output estimation by thermodilution during artificial ventilation was studied in anesthetized pigs at the right side of the heart. The estimates exhibited a cyclic modulation related to the ventilation. The amplitude of the modulation was independent of the level of positive end-expiratory pressure, ventilatory pattern and volemic loading of the animals. However, a non-constant phase relation existed between the ventilatory cycle and the modulation. Single observations at a fixed moment in the ventilatory cycle are therefore not appropriate for estimation of mean cardiac output nor for studying its relative changes. The averaging of estimates spread equally over the ventilatory cycle led to a much larger reduction in the deviation of the averages from the mean cardiac output than an averaging procedure of randomly selected estimates. The accuracy of estimation of mean cardiac output by two estimates equally spread in the ventilatory cycle was equal to the accuracy obtained by averaging five randomly selected estimates. Averaging four estimates, equally spread in the cycle, appeared to be the optimal procedure. For 89% of all averages an accuracy of 5% around the mean was obtained and for 99% an accuracy of +/- 10%.

摘要

在麻醉猪的心脏右侧,研究了人工通气期间通过热稀释法估计心输出量的可靠性。这些估计值呈现出与通气相关的周期性调制。调制幅度与呼气末正压水平、通气模式和动物的血容量负荷无关。然而,通气周期与调制之间存在非恒定的相位关系。因此,在通气周期的固定时刻进行单次观察,既不适用于估计平均心输出量,也不适用于研究其相对变化。与随机选择估计值的平均程序相比,将估计值平均分布在整个通气周期导致平均值与平均心输出量偏差的减小幅度要大得多。在通气周期中均匀分布的两个估计值对平均心输出量的估计准确性,与对五个随机选择估计值进行平均所获得的准确性相当。在周期中均匀分布的四个估计值进行平均似乎是最佳程序。对于所有平均值的89%,在平均值周围获得了5%的准确性,对于99%,获得了+/- 10%的准确性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验