Feasibility and repeatability of ocular biometry measured with IOLMaster 700 in a large population-based study.

机构信息

Institute for Medical Informatics, Statistics, and Epidemiology (IMISE), Leipzig University, Leipzig, Germany.

Medical Informatics Center - Dept. of Medical Data Science, Leipzig University Medical Centre, Leipzig, Germany.

出版信息

Ophthalmic Physiol Opt. 2023 Jul;43(4):860-873. doi: 10.1111/opo.13148. Epub 2023 Apr 21.

Abstract

PURPOSE

To evaluate the feasibility and repeatability of IOLMaster 700 biometry measurements in an adult population. Furthermore, to assess the value of the Quality Indicators (QIs) provided by the device.

METHOD

As part of the large population-based Leipzig Research Centre for Civilization Diseases (LIFE) Adult-Study, randomly selected participants from Leipzig, Germany were evaluated with the ZEISS IOLMaster 700. Age range was 26-85 years, with 53% of participants above 70 years of age. Axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT) and keratometry (K) were assessed in 1767 right eyes. Measurements were repeated twice and in a subset of 1331 eyes, three times. Measurement feasibility was evaluated for three levels; successful, with warnings and failed, using the inbuilt QIs. Repeatability was assessed as within-subject standard deviation (SD) and repeatability limits were calculated.

RESULTS

First measurement success rate for phakic eyes was over 99% for AL, CCT, ACD, over 98% for LT and over 97% for K. K had 16% eyes with warnings and the recommendation to repeat the measurement. Excluding the measurements with warnings resulted in a reduction of mean SD for AL from 48 to 4 μm and for mean K from 0.08 to 0.04 D. Repeatability for phakic eyes was 8 μm for AL, CCT, ACD and LT and 2.3 μm for CCT; 0.07 D and 0.12 D for mean K and delta K, respectively, for phakic cases without warnings (two measurements).

CONCLUSIONS

In our population-based sample, the IOLMaster 700 collected data for AL, CCT, ACD, LT and K from the vast majority of eyes. Considering the built-in QIs improved the measurement variability substantially. Repeatability measurements indicate that clinically meaningful changes can be detected reliably with this instrument.

摘要

目的

评估 IOLMaster 700 生物测量仪在成年人群中的可行性和可重复性。此外,评估设备提供的质量指标(QIs)的价值。

方法

作为大型基于人群的莱比锡文明疾病研究中心(LIFE)成人研究的一部分,从德国莱比锡随机选择参与者,使用蔡司 IOLMaster 700 进行评估。年龄范围为 26-85 岁,其中 53%的参与者年龄超过 70 岁。在 1767 只右眼评估了眼轴(AL)、中央角膜厚度(CCT)、前房深度(ACD)、晶状体厚度(LT)和角膜曲率(K)。测量重复两次,在 1331 只眼中的一个亚组中重复三次。使用内置 QIs,根据三个级别评估测量的可行性;成功、有警告和失败。评估重复性为个体内标准差(SD),并计算重复性限值。

结果

有晶状体眼的初次测量成功率,AL、CCT、ACD 的成功率超过 99%,LT 的成功率超过 98%,K 的成功率超过 97%。K 有 16%的眼睛有警告,并建议重复测量。排除有警告的测量结果,AL 的平均 SD 从 48μm 降至 4μm,平均 K 从 0.08D 降至 0.04D。有晶状体眼的重复性为 AL、CCT、ACD 和 LT 的 8μm,CCT 的 2.3μm;对于没有警告的有晶状体眼(两次测量),AL、CCT、ACD 和 LT 的平均 K 和 delta K 分别为 0.07D 和 0.12D。

结论

在我们的基于人群的样本中,IOLMaster 700 从绝大多数眼睛中收集了 AL、CCT、ACD、LT 和 K 的数据。考虑到内置的 QIs,大大改善了测量的可变性。重复性测量表明,该仪器可以可靠地检测到具有临床意义的变化。