Repeatability and reproducibility of optical biometry implemented in a new optical coherence tomographer and comparison with a optical low-coherence reflectometer.

机构信息

Hygeia Clinic, Gdańsk, Poland.

Stein Eye Institute, University of California, Los Angeles, USA; St. Mary's Eye Center, Santa Monica, California, USA.

出版信息

J Cataract Refract Surg. 2019 Nov;45(11):1619-1624. doi: 10.1016/j.jcrs.2019.07.002.

Abstract

PURPOSE

To examine the repeatability and reproducibility of biometry provided by a new anterior-posterior segment spectral-domain optical coherence tomography (SD-OCT) device (Revo NX) and compare it with that provided by a validated optical low-coherence reflectometry (OLCR) biometer (Lenstar LS 900).

SETTING

Hygeia Clinic, Gdańsk, Poland.

DESIGN

Prospective evaluation of diagnostic test.

METHODS

The axial length (AL), lens thickness (LT), central corneal thickness (CCT), and anterior chamber depth (ACD) (corneal epithelium to lens) were measured in eyes of volunteers. Two skilled operators obtained 3 measurements in random order with the SD-OCT device. Repeatability was calculated for each observer, while reproducibility was assessed by a change in observer. To compare agreement between the SD-OCT device and OLCR device, the first observer took 3 measurements with both devices.

RESULTS

The study evaluated 63 eyes of 63 volunteers. The mean AL measurement (+0.11 mm ± 0.02 [SD]) and mean ACD measurement (0.05 ± 0.04 mm) were significantly higher with the SD-OCT device than with the OLCR device; the CCT measurements did not differ between the devices. A very strong correlation was found for all parameters, while Bland-Altman analysis showed narrow 95% limits of agreement for AL (0.06 to 0.15 mm), ACD (-0.02 to 0.12 mm), LT (-0.23 to 0.17 mm), and CCT (-7.83 to 6.17 μm).

CONCLUSIONS

The new SD-OCT device provided repeatable and reproducible AL, ACD, LT, and CCT measurements. The results with the SD-OCT device showed strong correlations with those obtained with the OLCR device. However, the AL and ACD measurements of the 2 devices cannot be considered interchangeable.

摘要

目的

研究一种新型前节光谱域光相干断层扫描仪(SD-OCT)设备(Revo NX)提供的生物测量值的可重复性和再现性,并将其与经过验证的光学低相干反射仪(OLCR)生物测量仪(Lenstar LS 900)进行比较。

设置

波兰格但斯克 Hygia 诊所。

设计

前瞻性诊断测试评估。

方法

志愿者的眼轴长度(AL)、晶状体厚度(LT)、中央角膜厚度(CCT)和前房深度(ACD)(角膜上皮至晶状体)由 2 名熟练的操作人员使用 SD-OCT 设备进行 3 次随机测量。计算每位观察者的重复性,通过观察者的变化评估再现性。为了比较 SD-OCT 设备和 OLCR 设备之间的一致性,第一观察者使用两种设备各进行 3 次测量。

结果

该研究评估了 63 名志愿者的 63 只眼。与 OLCR 设备相比,SD-OCT 设备的平均 AL 测量值(+0.11 ± 0.02 [SD])和平均 ACD 测量值(0.05 ± 0.04mm)更高,而 CCT 测量值在两种设备之间没有差异。所有参数均显示出非常强的相关性,而 Bland-Altman 分析显示 AL(0.06 至 0.15mm)、ACD(-0.02 至 0.12mm)、LT(-0.23 至 0.17mm)和 CCT(-7.83 至 6.17μm)的 95%一致性限较窄。

结论

新型 SD-OCT 设备提供了可重复和可再现的 AL、ACD、LT 和 CCT 测量值。SD-OCT 设备的结果与 OLCR 设备的结果具有很强的相关性。然而,两种设备的 AL 和 ACD 测量值不能视为可互换。