Dupont G, Moriette G, Pajot N, Walti H, De Gamarra E, Relier J P
Arch Fr Pediatr. 1986 Feb;43(2):111-3.
Blood glucose levels were assessed in 49 neonates, using 2 blood glucose test strips. Results were compared to blood glucose levels. Correlations between values from the test strips and values from the laboratory were better using the Haemo-Glukotest (r = 0.91) than using the Dextrostix (r = 0.82). As compared to values from the laboratory, however, both glucose test strips gave higher values and did not properly identify all cases of hypoglycemia.