Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, East Melbourne, Australia.
Ophthalmology, Department of Surgery, The University of Melbourne, Melbourne, Australia.
Transl Vis Sci Technol. 2023 Jul 3;12(7):11. doi: 10.1167/tvst.12.7.11.
To examine whether systematic changes in visual sensitivity measurements on microperimetry occur over tests within the same session and whether these changes vary according to the level of visual sensitivity loss.
Eighty individuals with glaucoma or atrophic age-related macular degeneration underwent three microperimetry tests in one eye during one session using the 4-2 staircase strategy. Changes in mean sensitivity (MS) and pointwise sensitivity (PWS) between the first and second test pairs were examined, with PWS was examined separately based on its average value across the three tests in 6-dB bins. The coefficient of repeatability (CoR) for MS between each sequential test pair was also calculated.
There was a significant decline in MS from the first to second test (P = 0.001), but no significant difference in MS was seen between the second and third tests (P = 0.562). This significant decline in the first test pair was observed in locations with an average PWS of <6 dB or between 6 to 12 dB and between 12 to 18 dB (P < 0.001), but not for all other average PWS bins (P ≥ 0.337). The CoR of MS was significantly lower in the second compared to the first test pair (1.4 dB and 2.5 dB, respectively; P < 0.001).
The 4-2 staircase strategy conventionally used on microperimetry testing systematically underestimates visual sensitivity loss on the first test.
The consistency and accuracy of visual sensitivity measurements on microperimetry in clinical trials could be markedly improved by using estimates from an initial test to seed subsequent tests and excluding this first test from analyses.
研究在同一次检查中微视野计检查中视觉敏感性测量是否会发生系统性变化,以及这些变化是否因视觉敏感性损失程度而异。
80 名青光眼或萎缩性年龄相关性黄斑变性患者在同一次检查中使用 4-2 阶梯策略在一只眼中进行了三次微视野计检查。检查了第一对测试和第二对测试之间平均敏感度(MS)和逐点敏感度(PWS)的变化,并根据三次测试中 PWS 的平均值将其分别在 6-dB 间隔内进行检查。还计算了每对连续测试之间 MS 的可重复性系数(CoR)。
MS 从第一对测试到第二对测试显著下降(P=0.001),但第二对测试和第三对测试之间 MS 无显著差异(P=0.562)。在平均 PWS<6dB 或 6-12dB 之间以及 12-18dB 之间的位置观察到第一对测试中的显著下降(P<0.001),但在所有其他平均 PWS 间隔内未观察到这种情况(P≥0.337)。与第一对测试相比,第二对测试的 MS CoR 明显降低(分别为 1.4dB 和 2.5dB;P<0.001)。
微视野计检查中常规使用的 4-2 阶梯策略系统地低估了第一次测试中的视觉敏感性损失。