Sysmex DI-60 (Sysmex, Kobe, Japan) is a digital morphology (DM) analyzer widely used in clinical laboratories and supports body fluid (BF) applications. We evaluated analytical performance of DI-60 compared with XN-350 (Sysmex) and manual counting for BF cell differential counts.
A total of 213 BF samples were collected (47 cerebrospinal fluid [CSF], 80 pleural fluid, and 86 ascites samples). The analytical performance of DI-60 for BF cell differential counts was evaluated based on sensitivity, specificity, accuracy, and agreement. BF cell differential counts obtained by DI-60 were compared with those obtained by XN-350 and manual counting.
The overall sensitivity was high for neutrophils, lymphocytes, and macrophages (range, 83.1-99.4%). The overall specificity and overall accuracy were high for all cell types (range, 95.3-99.7% and 94.3-99.3%, respectively). The agreement between DI-60 pre-classification and verification was strong (κ = 0.89). The absolute mean differences between DI-60 verification and XN-350 ranged from 0.26 to 11.05, and differences between DI-60 verification and manual counting ranged from 0.01 to 4.76.
This is the first study to evaluate the performance of DI-60 compared with XN-350 and manual counting for BF cell differential counts. DI-60 showed reliable performance with CSF, pleural fluid, and ascites samples. For BF cell differential counts, DI-60 may be a better option than XN-350 and could be used for screening purposes in understaffed laboratories. To improve the hematology workflow for BF cell differential counting, the DM analyzer needs to be optimized by taking into account the laboratory situation and unmet needs, and the clinical laboratory needs to establish criteria for verification and manual slide review.
Sysmex DI-60(Sysmex,神户,日本)是一种广泛应用于临床实验室的数字形态学(DM)分析仪,支持体液(BF)应用。我们评估了 DI-60 与 XN-350(Sysmex)和手动计数在 BF 细胞分类计数方面的分析性能。
共收集了 213 例 BF 样本(47 例脑脊液[CSF]、80 例胸腔积液和 86 例腹水样本)。根据敏感性、特异性、准确性和一致性,评估 DI-60 对 BF 细胞分类计数的分析性能。比较 DI-60 与 XN-350 和手动计数获得的 BF 细胞分类计数。
中性粒细胞、淋巴细胞和巨噬细胞的总敏感性较高(范围为 83.1-99.4%)。所有细胞类型的总特异性和总准确性均较高(范围分别为 95.3-99.7%和 94.3-99.3%)。DI-60 预分类和验证之间的一致性较强(κ=0.89)。DI-60 验证与 XN-350 之间的绝对平均差异范围为 0.26-11.05,DI-60 验证与手动计数之间的差异范围为 0.01-4.76。
这是第一项评估 DI-60 与 XN-350 和手动计数在 BF 细胞分类计数方面性能的研究。DI-60 对 CSF、胸腔积液和腹水样本表现出可靠的性能。对于 BF 细胞分类计数,DI-60 可能是比 XN-350 更好的选择,并且可以在人员不足的实验室中用于筛选目的。为了改善 BF 细胞分类计数的血液学工作流程,需要根据实验室情况和未满足的需求优化 DM 分析仪,并由临床实验室建立验证和手动载玻片审查标准。