中央角膜厚度的测量。

The measurement of Central Corneal Thickness.

机构信息

"Carol Davila" University of Medicine and Pharmacy, Bucharest, Romania.

出版信息

Rom J Ophthalmol. 2023 Apr-Jun;67(2):168-174. doi: 10.22336/rjo.2023.29.

Abstract

We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, refractive surgery and keratoplasty. We measured the CCT using OCT, Topographer (TOPO), Ultrasonography Pachymeter (US), Specular Microscope (MS), and Non-contact Tonometer (TONO). In the analysis of the data collected from 59 patients we found the following mean values: US - 554.51 ± 29.849 μm, OCT - 548.73 ± 31.080 μm, TOPO - 553.76 ± 29.845 μm, MS - 564 ± 32.637 μm, and TONO - 538.9 ± 35.657 μm. Our results confirmed the strong correlation between techniques. OCT = Optical Coherence Tomography, CCT = Central Corneal Thickness, TOPO = Topographer, US = Ultrasonography Pachymeter, MS = Specular Microscope, TONO = Non-contact Tonometer.

摘要

我们旨在比较五种不同的角膜中央厚度测量设备。角膜中央厚度(CCT)是眼科学中的一个重要参数。它涉及到各种眼部疾病的管理,如青光眼、圆锥角膜、隐形眼镜佩戴、角膜营养不良、屈光手术和角膜移植。我们使用 OCT、 topography(TOPO)、超声角膜测厚仪(US)、共焦显微镜(MS)和非接触眼压计(TONO)测量 CCT。在对 59 名患者的数据进行分析后,我们发现以下平均值:US - 554.51 ± 29.849 μm,OCT - 548.73 ± 31.080 μm,TOPO - 553.76 ± 29.845 μm,MS - 564 ± 32.637 μm,TONO - 538.9 ± 35.657 μm。我们的结果证实了这些技术之间的强相关性。OCT = 光学相干断层扫描,CCT = 角膜中央厚度,TOPO = topography,US = 超声角膜测厚仪,MS = 共焦显微镜,TONO = 非接触眼压计。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/5c7e/10385715/0ec2e638360d/RomJOphthalmol-67-168-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验