Kim Chul, Song Jun Hyeong, Kim Seung Hyoun
Department of Rehabilitation Medicine, Inje University Sanggye Paik Hospital, Inje University College of Medicine, Seoul, Korea.
Ann Rehabil Med. 2023 Aug;47(4):261-271. doi: 10.5535/arm.23019. Epub 2023 Aug 4.
To assess the accuracy of recently commercialized wearable devices in heart rate (HR) measurement during cardiopulmonary exercise test (CPX) under gradual increase in exercise intensity, while wearable devices with HR monitors are reported to be less accurate in different exercise intensities.
CPX was performed for patients with coronary artery disease (CAD). Twelve lead electrocardiograph (ECG) was the gold standard and Apple watch 7 (AW7), Galaxy watch 4 (GW4) and Bio Patch Mobicare 200 (MC200) were applied for comparison. Paired absolute difference (PAD), mean absolute percentage error (MAPE) and intraclass correlation coefficient (ICC) were evaluated for each device.
Forty-four participants with CAD were included. All the devices showed MAPE under 2% and ICC above 0.9 in rest, exercise and recovery phases (MC200=0.999, GW4=0.997, AW7=0.998). When comparing exercise and recovery phase, PAD of MC200 and AW7 in recovery phase were significantly bigger than PAD of exercise phase (p<0.05). Although not significant, PAD of GW4 tended to be bigger in recovery phase, too. Also, when stratified by HR 20, ICC of all the devices were highest under HR of 100, and ICC decreased as HR increased. However, except for ICC of GW4 at HR above 160 (=0.867), all ICCs exceeded 0.9 indicating excellent accuracy.
The HR measurement of the devices validated in this study shows a high concordance with the ECG device, so CAD patients may benefit from the devices during high-intensity exercise under conditions where HR is measured reliably.
评估近期商业化的可穿戴设备在心肺运动试验(CPX)中运动强度逐渐增加时心率(HR)测量的准确性,而据报道带有心率监测器的可穿戴设备在不同运动强度下准确性较低。
对冠心病(CAD)患者进行CPX。以十二导联心电图(ECG)作为金标准,应用苹果手表7(AW7)、三星Galaxy手表4(GW4)和Bio Patch Mobicare 200(MC200)进行比较。对每个设备评估配对绝对差值(PAD)、平均绝对百分比误差(MAPE)和组内相关系数(ICC)。
纳入了44名CAD患者。在静息、运动和恢复阶段,所有设备的MAPE均低于2%,ICC均高于0.9(MC200 = 0.999,GW4 = 0.997,AW7 = 0.998)。在比较运动和恢复阶段时,MC200和AW7在恢复阶段的PAD显著大于运动阶段的PAD(p < 0.05)。虽然不显著,但GW4在恢复阶段的PAD也有增大趋势。此外,按心率分层时,所有设备的ICC在心率100以下时最高,且随着心率增加ICC下降。然而,除GW4在心率高于160时的ICC(= 0.867)外,所有ICC均超过0.9,表明准确性极佳。
本研究中验证的设备的心率测量与心电图设备显示出高度一致性,因此CAD患者在心率测量可靠的条件下进行高强度运动时可能会从这些设备中受益。