血压测量中的观察者误差。

Observer error in blood pressure measurement.

作者信息

Neufeld P D, Johnson D L

出版信息

CMAJ. 1986 Sep 15;135(6):633-7.

Abstract

This paper describes an experiment undertaken to determine observer error in measuring blood pressure by the auscultatory method. A microcomputer was used to display a simulated mercury manometer and play back tape-recorded Korotkoff sounds synchronized with the fall of the mercury column. Each observer's readings were entered into the computer, which displayed a histogram of all readings taken up to that point and thus showed the variation among observers. The procedure, which could easily be adapted for use in teaching, was used to test 311 observers drawn from physicians, nurses, medical students, nursing students and others at nine health care institutions in Ottawa. The results showed a strong bias for even-digit readings and standard deviations of roughly 5 to 6 mm Hg. The standard deviation for the systolic readings was somewhat smaller for the physicians as a group than for the nurses (3.5 v. 5.9 mm Hg). However, the standard deviations for the diastolic readings were roughly equal for these two groups (approximately 5.5 mm Hg).

摘要

本文描述了一项为确定听诊法测量血压时的观察者误差而进行的实验。使用一台微型计算机来显示模拟汞柱血压计,并回放与汞柱下降同步的录音柯氏音。每个观察者的读数都输入到计算机中,计算机显示截至该点所获取的所有读数的直方图,从而展示观察者之间的差异。该程序易于改编用于教学,被用于测试从渥太华九家医疗机构的医生、护士、医学生、护理学生及其他人员中抽取的311名观察者。结果显示,读数存在偏向偶数的强烈偏差,标准差约为5至6毫米汞柱。作为一个群体,医生的收缩压读数标准差比护士的略小(3.5对5.9毫米汞柱)。然而,这两组的舒张压读数标准差大致相等(约5.5毫米汞柱)。