Cummings Arthur B, Naughton Stephanie, Coen Ann-Marie, Brennan Elizabeth, Kelly Gabrielle E
The Wellington Eye Clinic, Dublin, D18 T8P3, Ireland.
Clin Ophthalmol. 2020 Dec 2;14:4209-4220. doi: 10.2147/OPTH.S278589. eCollection 2020.
To compare the accuracy of pre-operative corneal measurements obtained with four devices, and the refractive outcomes of two optical biometers.
Private practice.
Retrospective.
Data taken from biometric measurements on 299 consecutive eyes prior to cataract surgery were retrospectively analyzed using the Argos SS-Optical Biometer and the Lenstar LS900 PCI optical biometer. As part of the standard cataract surgery pre-operative exam, patients also underwent placido disk topography and Scheimpflug tomography. Keratometry, anterior chamber depth, corneal diameter, pupil diameter, central corneal thickness and axial length were all measured. The comparable measurements were compared. Finally, for those eyes where cataract surgery was performed, the post-operative refractive results were compared to the predictive results of the two biometers.
The SS-OCT Argos was able to measure all eyes, while five eyes could not be measured with the Lenstar LS900 PCI. Axial length measurements were performed only with the Argos and Lenstar devices. The eyes that could not be measured by the Lenstar LS900 PCI included dense grade IV nuclear sclerosis and large posterior subcapsular cataracts. In the primary endpoints, there was strong correlation between the Argos and the Lenstar devices in eyes with an axial length between 20 and 30 mm.
The predictive accuracies of the Argos Optical Biometer and Lenstar LS900 PCI are similar, except in medium and long eyes, in which the predictive accuracy of Argos SS-OCT biometry was higher. The Argos system was found easier to use by technicians when compared to the other biometry devices.
比较四种设备术前角膜测量的准确性以及两种光学生物测量仪的屈光结果。
私人诊所。
回顾性研究。
回顾性分析连续299只白内障手术前眼的生物测量数据,使用阿高斯SS-Optical生物测量仪和Lenstar LS900 PCI光学生物测量仪。作为标准白内障手术术前检查的一部分,患者还接受了Placido盘地形图和Scheimpflug断层扫描。测量了角膜曲率、前房深度、角膜直径、瞳孔直径、中央角膜厚度和眼轴长度。对可比较的测量结果进行了比较。最后,对于那些接受了白内障手术的眼睛,将术后屈光结果与两种生物测量仪的预测结果进行了比较。
阿高斯SS-OCT能够测量所有眼睛,而Lenstar LS900 PCI无法测量五只眼睛。眼轴长度测量仅使用阿高斯和Lenstar设备进行。Lenstar LS900 PCI无法测量的眼睛包括IV级核硬化严重和后囊下大白内障。在主要终点方面,眼轴长度在20至30毫米之间的眼睛中,阿高斯和Lenstar设备之间存在很强的相关性。
阿高斯光学生物测量仪和Lenstar LS900 PCI的预测准确性相似,但在中长眼方面,阿高斯SS-OCT生物测量的预测准确性更高。与其他生物测量设备相比,技术人员发现阿高斯系统更易于使用。