比较扫频源光学相干断层扫描生物测量仪和光学低相干反射测量生物测量仪的预测准确性。

Comparing Predictive Accuracy of a Swept Source Optical Coherence Tomography Biometer and an Optical Low Coherence Reflectometry Biometer.

作者信息

Blehm Clayton, Hall Brad

机构信息

Gainesville Eye Associates, Gainesville, GA, USA.

Sengi, Penniac, NB, Canada.

出版信息

Clin Ophthalmol. 2023 Jul 25;17:2125-2131. doi: 10.2147/OPTH.S421504. eCollection 2023.

Abstract

PURPOSE

To compare the refractive accuracy resulting from calculations based on measurements with a swept-source optical coherence tomography (SS-OCT) biometer compared to calculations based on measurements with an optical low coherence reflectometry (OLCR) biometer at one month postoperatively.

METHODS

This was a retrospective comparative non-interventional study of preoperative biometry and postoperative refraction and visual acuity of 200 eyes. All eyes had preoperative biometry with both the Argos (Movu, a Santec company) and Lenstar LS900 (Haag-Streit AG) devices. Data were collected for mean postoperative prediction error (directional and absolute), preoperative mean K, delta K (corneal astigmatism), axial length, and anterior chamber depth.

RESULTS

The mean directional prediction error was -0.15 ± 0.47 D for Argos and -0.31 ± 0.50 D for Lenstar LS900, and there was a statistically significant mean of the differences (0.16 ± 0.24 D; p < 0.001). The mean absolute prediction error was 0.35 ± 0.34 D for Argos and 0.42 ± 0.41 D for Lenstar LS900, and there was a statistically significant mean of the differences (-0.07 ± 0.24 D; p < 0.001). Neither the differences in directional prediction error nor the differences in absolute prediction error were clinically significant.

CONCLUSION

The directional and absolute prediction accuracies were statistically significant, but not clinically different between the Argos and Lenstar LS900 devices. In addition, differences between preoperative K, AL, and ACD measurements were not clinically significant.

摘要

目的

比较术后1个月时,基于扫频光学相干断层扫描(SS-OCT)生物测量仪测量结果的计算与基于光学低相干反射测量(OLCR)生物测量仪测量结果的计算所得到的屈光精度。

方法

这是一项对200只眼睛的术前生物测量、术后屈光和视力进行的回顾性比较非干预性研究。所有眼睛均使用阿高斯(Movu,山特公司)和Lenstar LS900(哈格施特赖特股份公司)设备进行术前生物测量。收集术后平均预测误差(方向和绝对)、术前平均角膜曲率、角膜散光差值(ΔK)、眼轴长度和前房深度的数据。

结果

阿高斯的平均方向预测误差为-0.15±0.47D,Lenstar LS900为-0.31±0.50D,差异均值具有统计学意义(0.16±0.24D;p<0.001)。阿高斯的平均绝对预测误差为0.35±0.34D,Lenstar LS900为0.42±0.41D,差异均值具有统计学意义(-0.07±0.24D;p<0.001)。方向预测误差的差异和绝对预测误差的差异在临床上均无显著意义。

结论

阿高斯和Lenstar LS900设备在方向和绝对预测精度上具有统计学意义,但在临床上无差异。此外,术前角膜曲率、眼轴长度和前房深度测量的差异在临床上也无显著意义。