J Cataract Refract Surg. 2022 Jan 1;48(1):16-25. doi: 10.1097/j.jcrs.0000000000000690.
To evaluate repeatability and agreement between various biometric parameters using 6 biometers based on different optical technologies.
University of Valencia, Spain.
Prospective, comparative case series.
150 eyes were measured using the Aladdin, AL-Scan, Argos, IOLMaster700, Lenstar LS900, and OA-2000 biometers. Keratometry (K1 and K2), J0 and J45, central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), axial length (AL), white to white (WTW), and pupil size (PS) were measured 5 times with each device. Intrasubject SD, coefficient of variability (CoV), coefficient of repeatability, intraclass correlation coefficient, and Bland -Altman graphs were analyzed.
CoV values were <0.30% for K1, K2, and AL and up to 1.61% for CCT, ACD, LT, and WTW. PS values were higher (from 4.2% to 7.68%). There was statistically significant differences between biometers for all parameters evaluated (P < .001), and these differences varied as a function of the parameter analyzed. The limit of agreement (LoA) width of some comparisons for K1 and the majority for K2 were >0.50 diopter. A similar pattern was found for J0/J45. For CCT, many comparisons showed LoA width values of >25 μm. The LoA width for ACD ranged from 0.366 mm to 0.175 mm and for LT was about 0.2 mm. AL showed a highest LoA width of 0.225 mm. The LoA width for WTW was, in most cases, about ≥0.50 mm. The LoA width for PS ranged from 1.578 mm to 3.541 mm.
The 6 biometers provided repeatable measurements for the different parameters analyzed. The LoA obtained for each comparison should be analyzed carefully to consider the interchangeability of these devices.
评估 6 种基于不同光学技术的生物测量仪在各种生物测量参数中的重复性和一致性。
西班牙瓦伦西亚大学。
前瞻性、比较病例系列。
使用 Aladdin、AL-Scan、Argos、IOLMaster700、Lenstar LS900 和 OA-2000 生物测量仪对 150 只眼睛进行了测量。使用每种设备对角膜曲率(K1 和 K2)、J0 和 J45、中央角膜厚度(CCT)、前房深度(ACD)、晶状体厚度(LT)、眼轴长度(AL)、白到白(WTW)和瞳孔大小(PS)进行了 5 次测量。分析了各设备的个体内标准差(SD)、变异系数(CoV)、可重复性系数、组内相关系数和 Bland -Altman 图。
K1、K2 和 AL 的 CoV 值<0.30%,而 CCT、ACD、LT 和 WTW 的 CoV 值高达 1.61%。PS 值较高(4.2%至 7.68%)。所有评估参数的生物测量仪之间均存在统计学差异(P<.001),这些差异随分析参数的不同而变化。一些 K1 和大部分 K2 比较的差异界限(LoA)宽度>0.50 屈光度。J0/J45 也出现了类似的模式。对于 CCT,许多比较显示的 LoA 宽度值>25μm。ACD 的 LoA 宽度范围为 0.366mm 至 0.175mm,LT 的 LoA 宽度约为 0.2mm。AL 的 LoA 宽度最高,为 0.225mm。WTW 的 LoA 宽度在大多数情况下约为≥0.50mm。PS 的 LoA 宽度范围为 1.578mm 至 3.541mm。
6 种生物测量仪为分析的不同参数提供了可重复的测量结果。对于每个比较,应仔细分析获得的 LoA,以考虑这些设备的可互换性。