Repeatability of automated measurements by a new anterior segment optical coherence tomographer and biometer and agreement with standard devices.

机构信息

IRCCS - Fondazione Bietti, Rome, Italy.

Stein Eye Institute, University of California, Los Angeles, CA, USA.

出版信息

Sci Rep. 2021 Jan 13;11(1):983. doi: 10.1038/s41598-020-79674-4.

Abstract

We assess repeatability of automatic measurements of a new anterior segment optical coherence tomographer and biometer (ANTERION) and their agreement with those provided by an anterior segment-optical coherence tomography device combined with Placido-disk corneal topography (MS-39) and a validated optical biometer (IOLMaster 500). A consecutive series of patients underwent three measurements with ANTERION and one with MS-39. A subgroup of patients underwent biometry also with IOLMaster 500. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. Paired t-test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices. Repeatability of ANTERION measurements was high, with ICC > 0.98 for all parameters except astigmatism (0.963); all parameters apart from those related to astigmatism revealed a COV < 1%. Repeatability of astigmatism improved when only eyes whose keratometric astigmatism was higher than 1.0 D were investigated. Most measurements by ANTERION and MS-39 showed good agreement. No significant differences were found between measurements by ANTERION and IOLMaster, but for corneal diameter. ANTERION revealed high repeatability of automatic measurements and good agreement with both MS-39 and IOLMaster for most parameters.

摘要

我们评估了一种新型眼前节光学相干断层扫描仪和生物测量仪(ANTERION)的自动测量的可重复性及其与结合了 Placido 盘角膜地形图(MS-39)的眼前节光学相干断层成像设备和经过验证的光学生物测量仪(IOLMaster 500)提供的测量值的一致性。连续系列患者接受了三次 ANTERION 测量和一次 MS-39 测量。一部分患者还接受了 IOLMaster 500 的生物测量。通过个体内标准差、变异系数(COV)和组内相关系数(ICC)来评估可重复性。使用 95%一致性界限来研究一致性。进行配对 t 检验和 Wilcoxon 配对检验比较不同设备的测量值。除了散光(0.963)之外,ANTERION 测量的所有参数的 ICC 均大于 0.98,表明 ANTERION 测量的可重复性高;除了与散光相关的参数外,所有参数的 COV 均小于 1%。当仅研究角膜散光高于 1.0 D 的眼睛时,散光的可重复性得到改善。ANTERION 和 MS-39 的大多数测量值显示出良好的一致性。ANTERION 与 IOLMaster 的测量值之间没有发现显著差异,但角膜直径除外。ANTERION 显示出自动测量的高可重复性,并与 MS-39 和 IOLMaster 的大多数参数具有良好的一致性。

相似文献

2
Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus.
J Refract Surg. 2024 Jul;40(7):e445-e452. doi: 10.3928/1081597X-20240514-03. Epub 2024 Jul 1.
5
Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.
Graefes Arch Clin Exp Ophthalmol. 2022 Jul;260(7):2271-2281. doi: 10.1007/s00417-022-05579-9. Epub 2022 Feb 16.
6
Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices.
Clin Exp Ophthalmol. 2024 Nov;52(8):800-810. doi: 10.1111/ceo.14423. Epub 2024 Jul 21.

引用本文的文献

1
Repeatability of corneal curvature with the MS-39 anterior segment tomographer in a cataractous population.
PLoS One. 2025 Jul 18;20(7):e0327863. doi: 10.1371/journal.pone.0327863. eCollection 2025.
3
Comparison of ocular biometry and refractive outcomes using two swept-source optical coherence tomography-based biometers.
PLoS One. 2024 Dec 31;19(12):e0316439. doi: 10.1371/journal.pone.0316439. eCollection 2024.
6
Repeatability and agreement of two ocular biometers with single and dual Scheimpflug cameras in keratoconus eyes.
Graefes Arch Clin Exp Ophthalmol. 2024 Nov;262(11):3701-3709. doi: 10.1007/s00417-024-06511-z. Epub 2024 Jul 24.
8
Agreements' profile of Scheimpflug-based optical biometer with gold standard partial coherence interferometry.
Int J Ophthalmol. 2023 Dec 18;16(12):2095-2104. doi: 10.18240/ijo.2023.12.24. eCollection 2023.
10

本文引用的文献

1
Ocular biometric repeatability using a new high-resolution swept-source optical coherence tomographer.
Expert Rev Med Devices. 2020 Jun;17(6):591-597. doi: 10.1080/17434440.2020.1772050. Epub 2020 Jun 11.
2
Comparison of anterior segment and lens biometric measurements in patients with cataract.
Graefes Arch Clin Exp Ophthalmol. 2020 Jan;258(1):137-146. doi: 10.1007/s00417-019-04482-0. Epub 2019 Oct 20.
3
Agreement of ocular biometry measurements between 2 biometers.
J Cataract Refract Surg. 2019 Aug;45(8):1130-1134. doi: 10.1016/j.jcrs.2019.03.016. Epub 2019 Jul 3.
4
Comparative analysis of 2 swept-source optical coherence tomography biometers.
J Cataract Refract Surg. 2019 Aug;45(8):1124-1129. doi: 10.1016/j.jcrs.2019.03.020. Epub 2019 Jun 4.
6
Anterior segment optical coherence tomography.
Prog Retin Eye Res. 2018 Sep;66:132-156. doi: 10.1016/j.preteyeres.2018.04.002. Epub 2018 Apr 7.
7
In vivo crystalline lens measurements with novel swept-source optical coherent tomography: an investigation on variability of measurement.
BMJ Open Ophthalmol. 2017 Mar 3;1(1):e000058. doi: 10.1136/bmjophth-2016-000058. eCollection 2017.
9
Intrasession repeatability of pachymetry measurements with RTVue XR 100 optical coherence tomography in normal cornea.
Saudi J Ophthalmol. 2017 Apr-Jun;31(2):65-68. doi: 10.1016/j.sjopt.2017.04.003. Epub 2017 Apr 9.
10
Total Corneal Astigmatism Measurements: Agreement Between 2 Rotating Scheimpflug Cameras.
Cornea. 2017 Apr;36(4):463-469. doi: 10.1097/ICO.0000000000001117.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验