Automated endometrial identification and volume calculation in normal uteri using a novel smart ERA technique.

机构信息

Department of Ultrasound, Peking University Shenzhen Hospital, No.1120, Lianhua Road, Futian District, Shenzhen, 518036, Guangdong, China.

出版信息

Sci Rep. 2024 Sep 4;14(1):20525. doi: 10.1038/s41598-024-71069-z.

Abstract

To evaluate the repeatability of a novel automated technique called Smart ERA (Smart Endometrial Receptivity Analysis) for the automated segmentation and volume calculation of the endometrium in patients with normal uteri,, and to compare the agreement of endometrial volume measurements between Smart ERA, the semi-automated Virtual Organ Computer-aided Analysis (VOCAL) technique and manual segmentation. This retrospective study evaluated endometrial volume measurement in infertile patients who underwent frozen-thawed embryo transfer (FET). Transvaginal three-dimensional ultrasound scans were performed using a Resona R9 ultrasound machine. Data was collected from patients between 2021 and 2022. Patients with normal uteri and optimal ultrasound images were included. Endometrial volumes were measured using Smart ERA, VOCAL at 15° rotation, and manual segmentation. Intra-observer repeatability and agreement between techniques were assessed using the intraclass correlation coefficient (ICC) and Bland-Altman analysis. A total of 407 female patients were evaluated (mean age 33.2 ± 4.7 years). The repeatability of Smart ERA showed an ICC of 0.983 (95% CI 0.984-0.991). The agreement between Smart ERA and the manual method, Smart ERA and VOCAL, and VOCAL and the manual method, as assessed by ICC, were 0.986 (95% CI 0.977-0.990), 0.943 (95% CI 0.934-0.963), and 0.951 (95% CI 0.918-0.969), respectively. The Smart ERA technique required approximately 3 s for endometrial volume calculation, while VOCAL took around 5 min and the manual segmentation method took approximately 50 min. The Smart-ERA software, which employs a novel three-dimensional segmentation algorithm, demonstrated excellent intra-observer repeatability and high agreement with both VOCAL and manual segmentation for endometrial volume measurement in women with normal uteri. However, these findings should be interpreted with caution, as the algorithm's performance may not be generalizable to populations with different uterine characteristic. Additionally, Smart ERA required significantly less time compared to VOCAL and manual segmentation.

摘要

评估一种名为 Smart ERA(智能子宫内膜容受性分析)的新型自动化技术在正常子宫患者中的子宫内膜自动分割和体积计算的可重复性,并比较 Smart ERA、半自动 Virtual Organ Computer-aided Analysis (VOCAL) 技术和手动分割的子宫内膜体积测量的一致性。这项回顾性研究评估了接受冷冻胚胎移植 (FET) 的不孕患者的子宫内膜体积测量。使用 Resona R9 超声机进行经阴道三维超声扫描。数据是在 2021 年至 2022 年期间从患者中收集的。纳入了具有正常子宫和最佳超声图像的患者。使用 Smart ERA、15°旋转的 VOCAL 和手动分割测量子宫内膜体积。使用组内相关系数 (ICC) 和 Bland-Altman 分析评估观察者内重复性和技术之间的一致性。共评估了 407 名女性患者(平均年龄 33.2±4.7 岁)。Smart ERA 的重复性显示 ICC 为 0.983(95%CI 0.984-0.991)。Smart ERA 与手动方法、Smart ERA 与 VOCAL 以及 VOCAL 与手动方法的一致性,根据 ICC 评估,分别为 0.986(95%CI 0.977-0.990)、0.943(95%CI 0.934-0.963)和 0.951(95%CI 0.918-0.969)。Smart ERA 技术计算子宫内膜体积大约需要 3 秒,而 VOCAL 大约需要 5 分钟,手动分割方法大约需要 50 分钟。Smart-ERA 软件采用了一种新颖的三维分割算法,在测量正常子宫女性的子宫内膜体积方面,显示出了出色的观察者内重复性和与 VOCAL 和手动分割的高度一致性。然而,应该谨慎解释这些发现,因为该算法的性能可能不适用于具有不同子宫特征的人群。此外,Smart ERA 与 VOCAL 和手动分割相比,所需时间明显更少。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6d47/11371911/480ee5034dd4/41598_2024_71069_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验