Department of Bioengineering, Instituto Superior Técnico, University of Lisbon, 1949-001 Lisbon, Portugal.
IT-Instituto de Telecomunicações, Instituto Superior Técnico, 1949-001 Lisbon, Portugal.
Sensors (Basel). 2024 Aug 30;24(17):5622. doi: 10.3390/s24175622.
The high cost and limited availability of home spirometers pose a significant barrier to effective respiratory disease management and monitoring. To address this challenge, this paper introduces a novel Venturi-based spirometer designed for home use, leveraging the Bernoulli principle. The device features a 3D-printed Venturi tube that narrows to create a pressure differential, which is measured by a differential pressure sensor and converted into airflow rate. The airflow is then integrated over time to calculate parameters such as the Forced Vital Capacity (FVC) and Forced Expiratory Volume in one second (FEV). The system also includes a bacterial filter for hygienic use and a circuit board for data acquisition and streaming. Evaluation with eight healthy individuals demonstrated excellent test-retest reliability, with intraclass correlation coefficients (ICCs) of 0.955 for FVC and 0.853 for FEV. Furthermore, when compared to standard Pulmonary Function Test (PFT) equipment, the spirometer exhibited strong correlation, with Pearson correlation coefficients of 0.992 for FVC and 0.968 for FEV, and high reliability, with ICCs of 0.987 for FVC and 0.907 for FEV. These findings suggest that the Venturi-based spirometer could significantly enhance access to spirometry at home. However, further large-scale validation and reliability studies are necessary to confirm its efficacy and reliability for widespread use.
家用肺活量计价格高昂且供应有限,这对有效的呼吸疾病管理和监测构成了重大障碍。为了解决这一挑战,本文介绍了一种新颖的基于文丘里管的肺活量计,专为家庭使用而设计,利用了伯努利原理。该设备采用 3D 打印的文丘里管,管内逐渐变窄以产生压力差,由差压传感器测量,并转换为气流率。然后通过积分计算出用力肺活量(FVC)和 1 秒用力呼气量(FEV)等参数。该系统还包括一个用于卫生用途的细菌过滤器和一个用于数据采集和传输的电路板。通过对 8 名健康个体进行评估,该设备表现出了出色的测试-重测可靠性,FVC 的组内相关系数(ICC)为 0.955,FEV 的 ICC 为 0.853。此外,与标准的肺功能测试(PFT)设备相比,该肺活量计表现出了很强的相关性,FVC 的皮尔逊相关系数为 0.992,FEV 的皮尔逊相关系数为 0.968,且具有很高的可靠性,FVC 的 ICC 为 0.987,FEV 的 ICC 为 0.907。这些发现表明,基于文丘里管的肺活量计可以显著提高家庭进行肺活量计检查的机会。然而,需要进一步进行大规模的验证和可靠性研究,以确认其在广泛使用中的有效性和可靠性。