Abstract

BACKGROUND

Spirofy™ is India's first portable, pneumotach flow-sensor-based digital spirometer developed to diagnose asthma and chronic obstructive pulmonary disease (COPD). In this study, we compared the performance of the Spirofy™ device with that of the Vitalograph Alpha Touch™ spirometer in measuring the lung capacities of healthy individuals, asthmatics, and COPD patients. We also assessed the inter-device variability between two Spirofy™ devices.

METHODS

In a randomized, three-way crossover, open-label study, we measured the differences in forced expiratory volume in the first second (FEV) and forced vital capacity (FVC) between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. A proportion of the FEV/FVC ratio distribution of < 0.7 was used to compare the diagnostic accuracies of the Spirofy™ with Vitalograph™ Alpha Touch™ spirometers.

RESULTS

Ninety subjects participated in this study. The mean ± SD FVC values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 2.60 ± 1.05 L, 2.64 ± 1.04 L, and 2.67 ± 1.04 L, respectively. The mean ± SD FEV values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 1.87 ± 0.92 (L), 1.88 ± 0.92 (L), and 1.93 ± 0.93 (L), respectively. A significant positive correlation was found between the FVC and FEV values recorded by Vitalograph Alpha Touch™, Spirofy™ 1, and Spirofy™ 2. As compared to Vitalograph Alpha Touch™, the Spirofy™ device showed good sensitivity (97%), specificity (90%), and overall accuracy (93.3%) at an FEV/FVC ratio < 0.7. No inter-device variability was observed between the two Spirofy™ devices.

CONCLUSION

Spirofy™ is a portable and easy-to-use device and is as accurate as the standard Vitalograph Alpha Touch™ spirometer for the diagnosis of COPD and asthma.

TRIAL REGISTRATION

CTRI/2021/09/036492 (Clinical Trials Registry - India).

摘要

背景

Spirofy™ 是印度首款便携式、基于气动流量传感器的数字肺活量计,旨在诊断哮喘和慢性阻塞性肺疾病(COPD)。在这项研究中,我们比较了 Spirofy™ 设备与 Vitalograph Alpha Touch™ 肺活量计在测量健康个体、哮喘患者和 COPD 患者肺容量方面的性能。我们还评估了两台 Spirofy™ 设备之间的设备间可变性。

方法

在一项随机、三向交叉、开放标签研究中,我们测量了 Spirofy™ 和 Vitalograph Alpha Touch™ 肺活量计之间第一秒用力呼气量(FEV)和用力肺活量(FVC)的差异。使用 < 0.7 的 FEV/FVC 比值分布来比较 Spirofy™ 与 Vitalograph™ Alpha Touch™ 肺活量计的诊断准确性。

结果

90 名受试者参加了这项研究。Spirofy™ 1、Spirofy™ 2 和 Vitalograph Alpha Touch™ 设备获得的平均 ± 标准差 FVC 值分别为 2.60 ± 1.05 L、2.64 ± 1.04 L 和 2.67 ± 1.04 L。Spirofy™ 1、Spirofy™ 2 和 Vitalograph Alpha Touch™ 设备获得的平均 ± 标准差 FEV 值分别为 1.87 ± 0.92(L)、1.88 ± 0.92(L)和 1.93 ± 0.93(L)。Vitalograph Alpha Touch™、Spirofy™ 1 和 Spirofy™ 2 记录的 FVC 和 FEV 值之间存在显著的正相关。与 Vitalograph Alpha Touch™ 相比,Spirofy™ 设备在 FEV/FVC 比值 < 0.7 时显示出良好的敏感性(97%)、特异性(90%)和总准确性(93.3%)。两台 Spirofy™ 设备之间没有观察到设备间可变性。

结论

Spirofy™ 是一种便携式、易于使用的设备,与标准的 Vitalograph Alpha Touch™ 肺活量计一样准确,可用于诊断 COPD 和哮喘。

试验注册

CTRI/2021/09/036492(印度临床试验注册处)。