Sensors (Basel). 2024 Oct 1;24(19):6379. doi: 10.3390/s24196379.
(1) Background: Wearable sensors support healthcare professionals in clinical decision-making by measuring vital parameters such as heart rate (HR), respiration rate (RR), and blood oxygenation saturation (SpO). This study assessed the validity and reliability of two types of wearable sensors, based on electrocardiogram or photoplethysmography, compared with continuous monitoring of patients recovering from trauma surgery at the postanesthesia care unit. (2) Methods: In a prospective observational study, HR, RR, SpO, and temperature of patients were simultaneously recorded with the VitalPatch and Radius PPG and compared with reference monitoring. Outcome measures were formulated as correlation coefficient for validity and mean difference with 95% limits of agreement for reliability for four random data pairs and 30-min pairs per vital sign per patient. (3) Results: Included were 60 patients. Correlation coefficients for VitalPatch were 0.57 to 0.85 for HR and 0.08 to 0.16 for RR, and for Radius PPG, correlation coefficients were 0.60 to 0.83 for HR, 0.20 to 0.12 for RR, and 0.57 to 0.61 for SpO. Both sensors presented mean differences within the cutoff values of acceptable difference. (4) Conclusions: Moderate to strong correlations for HR and SpO were demonstrated. Although mean differences were within acceptable cutoff values for all vital signs, only limits of agreement for HR measured by electrocardiography were considered clinically acceptable.
(1)背景:可穿戴传感器通过测量心率(HR)、呼吸频率(RR)和血氧饱和度(SpO2)等生命参数,为医疗专业人员的临床决策提供支持。本研究评估了两种基于心电图或光电容积脉搏波的可穿戴传感器与术后恢复患者在麻醉后监护病房进行连续监测相比的有效性和可靠性。(2)方法:在一项前瞻性观察研究中,同时使用 VitalPatch 和 Radius PPG 记录患者的 HR、RR、SpO2 和体温,并与参考监测进行比较。制定了有效性的相关系数和可靠性的均值差异及其 95%一致性区间的测量指标,用于四个随机数据对和每个患者每体征 30 分钟的 30 对数据。(3)结果:共纳入 60 例患者。VitalPatch 的相关系数为 HR 0.57 至 0.85 和 RR 0.08 至 0.16,Radius PPG 的相关系数为 HR 0.60 至 0.83、RR 0.20 至 0.12 和 SpO2 0.57 至 0.61。两种传感器的所有生命体征的均值差异均在可接受差异的临界值范围内。(4)结论:证明了 HR 和 SpO2 的相关性为中度至高度。尽管所有生命体征的均值差异均在可接受的临界值范围内,但只有心电图测量的 HR 限制区间被认为具有临床可接受性。