Gaeth Catharina, Duarte Jamila, Rodriguez Alvaro, Powers Amber, Stone Randolph
Combat Wound Care, United States Army Institute of Surgical Research, 3698 Chambers Pass, JBSA Fort Sam Houston, Houston, TX 78234, USA.
Diagnostics (Basel). 2024 Nov 23;14(23):2641. doi: 10.3390/diagnostics14232641.
BACKGROUND/OBJECTIVES: Blood lactate concentration is often used to assess systemic hypoperfusion, tissue hypoxia, and sepsis in trauma patients and serves as a prognostic indicator and marker of response to therapy. Point-of-care (POC) devices provide rapid lactate measurements with a single drop of blood. In this study, lactate values from whole blood, measured with two POC devices, Abbott i-STAT and the Nova Biomedical Lactate (LA) Plus™ meter, are compared.
An observational analysis of 760 blood samples from four preclinical trauma animal models was performed. Lactate was measured simultaneously with two POC devices (i-STAT and the Lactate Plus™ meter). The evaluation in the form of a Bland-Altman diagram showed the agreement of the tests.
The Spearman correlation was 0.9180 (95% CI 0.9056 to 0.9288) for i-STAT vs. the LA Plus™ meter. Both tests showed a significant increase in lactate after injury ( < 0.05). The i-STAT showed a small bias compared to the LA Plus™ meter (-0.0227 ± 0.4542 with 95% limits of agreement from -0.09129 to 0.8675). The LA Plus™ meter showed slightly higher values overall (0.0227 mmol/L), but the i-STAT showed higher results at lactate levels above 2.5 mmol/L.
The observed similarity in performance between the two devices suggests that both are capable of reliably measuring lactate levels in a trauma situation. The LA Plus™ meter provides fast results with only one drop of blood. This study supports the prehospital use of POC devices.
背景/目的:血乳酸浓度常用于评估创伤患者的全身低灌注、组织缺氧和脓毒症,是一种预后指标及治疗反应的标志物。即时检验(POC)设备可通过一滴血快速检测乳酸水平。在本研究中,比较了使用两种POC设备(雅培i-STAT和诺华生物医学乳酸(LA)Plus™血糖仪)测量全血乳酸值的情况。
对来自四个临床前创伤动物模型的760份血样进行了观察性分析。同时使用两种POC设备(i-STAT和乳酸Plus™血糖仪)测量乳酸。以Bland-Altman图的形式进行评估,显示了测试结果的一致性。
i-STAT与LA Plus™血糖仪的Spearman相关性为0.9180(95%CI 0.9056至0.9288)。两种测试均显示受伤后乳酸显著升高(<0.05)。与LA Plus™血糖仪相比,i-STAT显示出较小的偏差(-0.0227±0.4542,95%一致性界限为-0.09129至0.8675)。LA Plus™血糖仪总体显示的值略高(0.0227 mmol/L),但i-STAT在乳酸水平高于2.5 mmol/L时显示出更高的结果。
观察到的两种设备性能的相似性表明,两者都能够在创伤情况下可靠地测量乳酸水平。LA Plus™血糖仪仅需一滴血就能快速得出结果。本研究支持POC设备在院前的使用。