在评估角膜移植术后圆锥角膜眼方面,Scheimpflug相机与光学相干断层扫描的比较。

Comparison of a Scheimpflug Camera and Optical Coherence Tomography in Evaluating Keratoconic Eyes Post Keratoplasty.

作者信息

Gadamer Anna Maria, Miklaszewski Piotr, Janiszewska-Bil Dominika, Lyssek-Boroń Anita, Dobrowolski Dariusz, Wylęgała Edward, Grabarek Beniamin Oskar, Krysik Katarzyna

机构信息

Department of Ophthalmology, St. Barbara Hospital, Trauma Centre, 41-200 Sosnowiec, Poland.

Department of Ophthalmology, Faculty of Medicine, Academy of Silesia, 40-555 Katowice, Poland.

出版信息

J Clin Med. 2025 Jan 3;14(1):238. doi: 10.3390/jcm14010238.

Abstract

The aim of this retrospective study was to compare corneal parameters and compliance using a Pentacam HR-Scheimpflug (Pentacam HR) and a swept-source OCT Casia (Casia) in keratoconus (KC) patients post penetrating keratoplasty (PKP) and KC patients without PKP, as well as a control group. Pachymetry measurements were also analyzed using a spectral domain OCT Solix (OCT Solix), Pentacam HR, and Casia. The study included 71 patients (136 keratoconic eyes; group A), 86 eyes with KC post-PKP (group B), 50 eyes with KC without PKP (group C), and 52 control participants (104 eyes). All participants were adults, Polish Caucasian, and met specific inclusion criteria. Patients with ophthalmological or systemic diseases, cognitive impairment, or pregnancy were excluded. Corneal parameters were measured using two devices (Casia and Pentacam HR), while pachymetry was assessed with three devices (Casia, Pentacam HR, and OCT Solix), with the inter-device agreement and group differences analyzed. Significant differences ( < 0.05) were found across all groups. The post-PKP KC eyes showed significant differences in all front parameters and K2 and Astig. back, while the non-PKP KC eyes showed differences in the K1 back ( = 0.025). The controls displayed differences in all parameters except front astigmatism ( = 0.61). The Pentacam HR overestimated the thinnest corneal thickness (TCT) compared to the OCT Casia across groups. The inter-device agreement was excellent for the anterior parameters (ICC > 0.9) but good for the posterior parameters and TCT. This study highlights significant variability in corneal and pachymetry measurements across devices, with OCT Casia providing more consistent and clinically reliable results than Pentacam HR. Clinicians should exercise caution when using these devices interchangeably, particularly for posterior parameters and TCT.

摘要

这项回顾性研究的目的是,在圆锥角膜(KC)穿透性角膜移植术(PKP)后患者、未行PKP的KC患者以及一个对照组中,比较使用Pentacam HR-Scheimpflug系统(Pentacam HR)和扫频源光学相干断层扫描Casia(Casia)测量的角膜参数及测量的依从性。还使用了光谱域光学相干断层扫描Solix(OCT Solix)、Pentacam HR和Casia分析角膜厚度测量值。该研究纳入了71例患者(136只圆锥角膜眼;A组)、86只PKP术后KC眼(B组)、50只未行PKP的KC眼(C组)以及52名对照参与者(104只眼)。所有参与者均为成年波兰白种人,且符合特定纳入标准。排除患有眼科或全身性疾病、认知障碍或妊娠的患者。使用两种设备(Casia和Pentacam HR)测量角膜参数,同时使用三种设备(Casia、Pentacam HR和OCT Solix)评估角膜厚度,并分析设备间的一致性和组间差异。所有组间均发现有显著差异(<0.05)。PKP术后KC眼在所有前表面参数、K2和后表面散光方面均有显著差异,而未行PKP的KC眼在K1后表面有差异(=0.025)。对照组除前表面散光外(=0.61),所有参数均有差异。与Casia相比,Pentacam HR在所有组中均高估了最薄角膜厚度(TCT)。设备间在前表面参数方面一致性极佳(组内相关系数>0.9),但在后表面参数和TCT方面一致性良好。本研究强调了不同设备在角膜和角膜厚度测量方面存在显著差异,与Pentacam HR相比,Casia提供的结果更一致且临床可靠性更高。临床医生在交替使用这些设备时应谨慎,尤其是在后表面参数和TCT方面。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/1b19/11722208/6df1a205a7ab/jcm-14-00238-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验