联网胰岛素笔系统的剂量准确性和可靠性

Dose Accuracy and Reliability of a Connected Insulin Pen System.

作者信息

Backfish Matthew, Kovalchick Kimberly, Albert Robert, Rosilio Myriam, Chigutsa Farai, Liao Birong

机构信息

Eli Lilly and Company, Indianapolis, IN, USA.

出版信息

J Diabetes Sci Technol. 2025 Jan 29:19322968241301760. doi: 10.1177/19322968241301760.

Abstract

For people with diabetes on insulin injection therapy, insulin pen dose accuracy and reliability are key features. Dose accuracy of the Tempo Pen with attached Tempo Smart Button [Module] (the system) was tested for three doses under standard atmospheric conditions. Reliability and the ability of the Module to detect, store, and transmit dose-related data were also tested. The system met the International Organization for Standardization 11608-1:2014 requirements for dose accuracy at all doses. Mean (standard deviation) doses were 0.010 mL (0.002), 0.299 mL (0.002), and 0.597 mL (0.004) for the 0.010-mL, 0.300-mL, and 0.600-mL doses, respectively. The Module also met requirements for data transfer after every injection. The system accurately delivered doses and reliably captured and stored insulin dosing information.

摘要

对于接受胰岛素注射治疗的糖尿病患者而言,胰岛素笔的剂量准确性和可靠性是关键特性。配备了Tempo智能按钮[模块](该系统)的Tempo笔的剂量准确性在标准大气条件下针对三种剂量进行了测试。还测试了该模块检测、存储和传输剂量相关数据的可靠性及能力。该系统在所有剂量下均符合国际标准化组织11608-1:2014对剂量准确性的要求。0.010 mL、0.300 mL和0.600 mL剂量的平均(标准差)剂量分别为0.010 mL(0.002)、0.299 mL(0.002)和0.597 mL(0.004)。该模块在每次注射后也符合数据传输要求。该系统能够准确输送剂量,并可靠地捕获和存储胰岛素给药信息。

相似文献

1
Dose Accuracy and Reliability of a Connected Insulin Pen System.
J Diabetes Sci Technol. 2025 Jan 29:19322968241301760. doi: 10.1177/19322968241301760.
2
Analysis of "Dose Accuracy and Reliability of a Connected Insulin Pen System".
J Diabetes Sci Technol. 2025 Jan 25:19322968251314522. doi: 10.1177/19322968251314522.
4
Dose Accuracy, Injection Force, and Usability Assessment of a New Half-Unit, Prefilled Insulin Pen.
J Diabetes Sci Technol. 2018 Mar;12(2):364-372. doi: 10.1177/1932296817736316. Epub 2017 Oct 30.
5
An assessment of dose accuracy and injection force of a novel prefilled insulin pen: comparison with a widely used prefilled insulin pen.
Expert Opin Drug Deliv. 2011 Oct;8(10):1271-6. doi: 10.1517/17425247.2011.615308. Epub 2011 Sep 2.
6
Expert Panel Recommendations for a Standardized Ambulatory Glucose Profile Report for Connected Insulin Pens.
Diabetes Technol Ther. 2024 Nov;26(11):814-822. doi: 10.1089/dia.2024.0107. Epub 2024 Jun 10.
8
Clinical, Economic, and Patient-Reported Benefits of Connected Insulin Pen Systems: A Systematic Literature Review.
Adv Ther. 2023 May;40(5):2015-2037. doi: 10.1007/s12325-023-02478-1. Epub 2023 Mar 16.
9
Functional Evaluation of the Reusable JuniorSTAR® Half-Unit Insulin Pen.
J Diabetes Sci Technol. 2015 May;9(3):625-31. doi: 10.1177/1932296815569246. Epub 2015 Jan 28.
10
Dose accuracy of the ClikSTAR, NovoPen 4, and Luxura insulin pens: results of laboratory and field studies.
J Diabetes Sci Technol. 2011 Sep 1;5(5):1179-84. doi: 10.1177/193229681100500522.

引用本文的文献

1
Understanding Suboptimal Insulin Use in Type 1 and Type 2 Diabetes: A Cross-Sectional Survey of People with Diabetes.
Patient Prefer Adherence. 2025 May 31;19:1625-1638. doi: 10.2147/PPA.S511332. eCollection 2025.

本文引用的文献

1
9. Pharmacologic Approaches to Glycemic Treatment: .
Diabetes Care. 2020 Jan;43(Suppl 1):S98-S110. doi: 10.2337/dc20-S009.
4
Mobile health in the management of type 1 diabetes: a systematic review and meta-analysis.
BMC Endocr Disord. 2019 Feb 13;19(1):21. doi: 10.1186/s12902-019-0347-6.
6
Rapid Evidence Review of Mobile Applications for Self-management of Diabetes.
J Gen Intern Med. 2018 Jul;33(7):1167-1176. doi: 10.1007/s11606-018-4410-1. Epub 2018 May 8.
9
Adherence of self-monitoring of blood glucose in persons with type 1 diabetes in Sweden.
BMJ Open Diabetes Res Care. 2017 Apr 6;5(1):e000342. doi: 10.1136/bmjdrc-2016-000342. eCollection 2017.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验