Accuracy of radiation field alignment in clinical practice.

作者信息

Rabinowitz I, Broomberg J, Goitein M, McCarthy K, Leong J

出版信息

Int J Radiat Oncol Biol Phys. 1985 Oct;11(10):1857-67. doi: 10.1016/0360-3016(85)90046-x.

Abstract

We present an analysis of simulator and portal films of 71 patients. Twenty-five were analyzed retrospectively, 39 prospectively, but without changing routine filming practice, and 7 had daily portal films taken. Treatment-to-treatment variations in anatomy with respect to the field were determined by comparing sequential portal films. The standard deviation of the variations was approximately normally distributed with an average value of 3 mm independent of site and field shaping technique. Discrepancies between the portal and simulator films were greater and depended on the site of treatment. The mean worst-case discrepancy averaged over all sites was 7.7 mm; the lowest value was 3.5 mm in the head and neck region; the highest value was 9.2 mm in the thorax.

摘要

我们对71例患者的模拟定位片和射野片进行了分析。其中25例为回顾性分析,39例为前瞻性分析,但未改变常规拍摄方法,另有7例进行了每日射野片拍摄。通过比较连续的射野片来确定不同治疗间靶区解剖结构的变化。这些变化的标准差大致呈正态分布,平均值为3毫米,与部位和射野塑形技术无关。射野片与模拟定位片之间的差异更大,且取决于治疗部位。所有部位的平均最坏情况差异为7.7毫米;头颈部最低值为3.5毫米;胸部最高值为9.2毫米。