Moore J G, Christian P E, Taylor A T, Alazraki N
J Nucl Med. 1985 Oct;26(10):1206-10.
Anteriorly acquired and geometric mean corrected gastric emptying curves of solids and liquid isotopic-labeled meals were compared in 37 subjects given 61 meals of three different sizes. Anterior data alone consistently and significantly underestimated solid-phase gastric emptying rates with all meal sizes when compared to geometric mean acquired data. However, with liquids there were only slight differences between anterior and anterior and posterior geometric mean corrected emptying-rates. The difference probably reflects greater attenuation of the 140 kev photon of 99mTc compared to the 247 keV photon of 111In. With anterior data alone, an apparent early delay in emptying of solids was present with all meal sizes and the resultant emptying curves were nonlinear in shape. Geometric mean correction resulted in the linearization of the solid-phase emptying curves and essentially eliminated the apparent delay in emptying or lag phase noted with the anterior data alone. Based on our results, geometric mean correction techniques are necessary for accurate assessment of radioisotopic-labeled solid meals.
在37名受试者中,给予61份三种不同大小的餐食,比较了固相和液相同位素标记餐食的前位采集和几何平均校正胃排空曲线。与几何平均采集数据相比,仅前位数据始终显著低估了所有餐食大小的固相胃排空率。然而,对于液体,前位和前后位几何平均校正排空率之间只有细微差异。这种差异可能反映了与111In的247 keV光子相比,99mTc的140 keV光子的衰减更大。仅使用前位数据时,所有餐食大小的固相排空均出现明显的早期延迟,并且所得排空曲线形状呈非线性。几何平均校正导致固相排空曲线线性化,并基本消除了仅使用前位数据时所观察到的明显排空延迟或滞后阶段。基于我们的结果,几何平均校正技术对于准确评估放射性同位素标记的固体餐食是必要的。