Pilossoff V, Schöber J G, Peters D, Bühlmeyer K
Eur J Pediatr. 1985 Nov;144(4):324-30. doi: 10.1007/BF00441772.
The accuracy of indirect oscillometric blood pressure measurements has been evaluated in 32 infants up to 11 months of age undergoing heart surgery. In a number of 1029 simultaneous measurements the indirect blood pressure was compared with the direct value obtained from a radial artery catheter. Cuffs of different sizes were applied. The main results were as follows: Regarding the ratio of cuff width to arm circumference, the best correlation between oscillometric and direct blood pressure measurements was obtained with a ratio of 0.38-0.41. The value of indirect blood pressure measurements depends on the absolute height of the blood pressure. In low blood pressure there is a tendency to underestimate and in high blood pressure there is a tendency to overestimate by the oscillometric blood pressure measurement. By applying an appropriate cuff size and by using our diagrams it should be possible to derive a direct value for the blood pressure on the basis of indirect oscillometric blood pressure measurements.
对32名年龄在11个月以下接受心脏手术的婴儿进行了间接示波法血压测量准确性的评估。在1029次同步测量中,将间接血压与通过桡动脉导管获得的直接血压值进行了比较。使用了不同尺寸的袖带。主要结果如下:关于袖带宽度与上臂周长的比例,示波法与直接血压测量之间的最佳相关性出现在比例为0.38 - 0.41时。间接血压测量值取决于血压的绝对高度。在低血压时,示波法血压测量有低估的趋势,在高血压时,有高估的趋势。通过应用合适的袖带尺寸并使用我们的图表,应该能够根据间接示波法血压测量得出直接血压值。