Pezzlo M T, Wetkowski M A, Peterson E M, de la Maza L M
J Clin Microbiol. 1983 Sep;18(3):697-701. doi: 10.1128/jcm.18.3.697-701.1983.
A study was conducted to evaluate the ability of a urine filtration system (Bac-T-Screen, Marion Laboratories, Inc., Kansas City, Mo.) to detect negative urine cultures within 2 min. A total of 1,000 urine specimens were tested with the Bac-T-Screen and compared with a standard semiquantitative culture plate method and the Autobac system (General Diagnostics, Warner-Lambert Co., Morris Plains, N.J.). Of the 1,000 clean voided urine specimens tested, 246 specimens had colony counts greater than or equal to 10(5) CFU/ml by the culture plate method. Of these, the Bac-T-Screen detected 65.4% (161 of 246), and the Autobac detected 63.0% (155 of 246). When pure cultures of diphtheroids, lactobacilli, and viridans streptococci other than group D and cultures containing multiple organisms were considered to be contaminants and, therefore, were excluded, there were 106 pure cultures of probable pathogens of which the Bac-T-Screen detected 76.4% (81 of 106) and the Autobac detected 90.6% (96 of 106). Some 133 specimens were uninterpretable with the Bac-T-Screen because 36 clogged the filter and 97 left a residual pigment on the filter. A majority of those clogging the filter (69.4%) had positive plate counts, whereas the majority of the pigmented urines had negative plate counts. Of those urine specimens tested. 754 were negative by the culture plate method. The false-positive rates for Bac-T-Screen and Autobac were 16.2 and 5.8%, respectively. As a urine screen, the Bac-T-Screen has a negative predictive value comparable to the Autobac system and has the advantage of being a 2-min test.
进行了一项研究,以评估尿液过滤系统(Bac-T-Screen,马里恩实验室公司,密苏里州堪萨斯城)在2分钟内检测阴性尿培养物的能力。总共1000份尿液标本用Bac-T-Screen进行检测,并与标准半定量培养平板法和自动细菌鉴定系统(通用诊断公司,华纳-兰伯特公司,新泽西州莫里斯平原)进行比较。在检测的1000份清洁中段尿标本中,通过培养平板法有246份标本的菌落计数大于或等于10⁵CFU/ml。其中,Bac-T-Screen检测出65.4%(246份中的161份),自动细菌鉴定系统检测出63.0%(246份中的155份)。当将类白喉杆菌、乳酸杆菌、非D群绿色链球菌的纯培养物以及含有多种微生物的培养物视为污染物并因此排除时,有106份可能病原体的纯培养物,其中Bac-T-Screen检测出76.4%(106份中的81份),自动细菌鉴定系统检测出90.6%(106份中的96份)。约133份标本用Bac-T-Screen无法解读,因为36份堵塞了过滤器,97份在过滤器上留下了残余色素。堵塞过滤器的标本大多数(69.4%)平板计数为阳性,而大多数有色素的尿液平板计数为阴性。在那些检测的尿液标本中,通过培养平板法754份为阴性。Bac-T-Screen和自动细菌鉴定系统的假阳性率分别为16.2%和5.8%。作为一种尿液筛查方法,Bac-T-Screen的阴性预测值与自动细菌鉴定系统相当,并且具有2分钟即可完成检测的优势。