静态和动态视野测试。正常志愿者的可重复性。

Static and kinetic visual field testing. Reproducibility in normal volunteers.

作者信息

Parrish R K, Schiffman J, Anderson D R

出版信息

Arch Ophthalmol. 1984 Oct;102(10):1497-502. doi: 10.1001/archopht.1984.01040031217021.

Abstract

The SD of static visual threshold ranges from 0.12 log units near fixation to 0.41 log units 30 degrees from fixation. Among 60 points tested, most persons have at least one point that tests 0.5 log units less sensitive than the adjacent points, simulating a scotoma 0.5 log units deep. With automated kinetic perimetry, the SD of isopter position is 2.6 degrees to 5.5 degrees, with the greatest variation appearing in the temporal field where the slope is flatter and the isopters are more peripheral. More than 10% of the points show an inward deviation of 5 degrees or more compared with their neighboring points. Retesting of the deviant points is required to distinguish an artifact caused by variable responsiveness from true localized defects. Judgments about progression also require that the accuracy of the isopter position or threshold determination be taken into account.

摘要

静态视觉阈值的标准差范围从注视点附近的0.12对数单位到距注视点30度处的0.41对数单位。在测试的60个点中,大多数人至少有一个点的测试敏感度比相邻点低0.5对数单位,模拟出一个深度为0.5对数单位的暗点。使用自动动态视野检查法时,等视线位置的标准差为2.6度至5.5度,最大变化出现在颞侧视野,该区域斜率较平缓且等视线更靠周边。超过10%的点与其相邻点相比向内偏差5度或更多。需要对偏差点进行重新测试,以区分由反应性变化引起的伪差和真正的局部缺陷。关于病情进展的判断还需要考虑等视线位置或阈值测定的准确性。