Keyzer J J, Oosting E, Wolthers B G, Muskiet F A, Hindriks F R, Van der Slik W
Pharm Weekbl Sci. 1983 Oct 21;5(5):248-51. doi: 10.1007/BF02332953.
The contamination caused by rubber stoppers of collection tubes preceding the determination of zinc in plasma and serum was investigated. The use of evacuated blood collection tubes (EBC-tubes), normally used for the collection of serum samples, could give rise to an average artificial increase of normal serum zinc concentrations of about 250%. Heparin containing EBC-tubes, used for the collection of plasma, and plain EBC-tubes of an experimental production appeared to be suitable for correct zinc determinations. There was no difference between the serum and plasma zinc levels of fifty volunteers.