用于锌分析的新生儿血样处理技术。

Technique of handling neonatal blood samples for zinc analysis.

作者信息

Michie D D, Bell N H, Wirth F H

出版信息

Am J Med Technol. 1976 Nov;42(11):424-7.

PMID:984078
Abstract

Short segments of Chemfluor flexible Teflon tubing (2.38 mm ID) previously coated with sodium heparin (1,000 units/ml) serve as ideal containers in which blood from neonates can be collected for zinc analysis. The ends of the tube are plugged with Teflon TFE round beading so that the sample comes in contact only with Teflon and the risk of contamination is minimized. The heparin coating facilitates easy collection of the sample and separation of plasma by centrifugation. It does not contribute zinc to the specimen.

摘要

预先涂有肝素钠(1000单位/毫升)的短段Chemfluor柔性聚四氟乙烯管(内径2.38毫米)是收集新生儿血液进行锌分析的理想容器。管子两端用聚四氟乙烯TFE圆珠堵塞,使样本仅与聚四氟乙烯接触,从而将污染风险降至最低。肝素涂层便于样本的采集以及通过离心分离血浆。它不会向样本中添加锌。