Konishi E, Takahashi J
J Clin Microbiol. 1983 Feb;17(2):225-31. doi: 10.1128/jcm.17.2.225-231.1983.
An enzyme-linked immunosorbent assay with polycarbonate-coated iron beads as the solid phase and with magnetic processing devices was evaluated for the quantitation of antibodies to Toxoplasma gondii in human serum samples. Under the parameters and other basic conditions determined in this study, the assay was highly reproducible: coefficients of variation for the absorbance values obtained with the positive serum were 2.42% in same-day tests and 3.75% in day-to-day tests. Significant correlations were observed between the present assay system and other conventional serological tests: correlation coefficients were 0.960 with the dye test and 0.929 with the latex agglutination test. Statistical analysis based on the frequency distribution of absorbance values for dye-test-positive and dye-test-negative serum samples gave feasible border lines for distinguishing between positive and doubtful samples (0.357) and between doubtful and negative samples (0.266). Under this diagnostic criterion, the results of our assay system agreed remarkably well with those obtained by the dye test and the latex agglutination test, with consistencies of 94.9 and 93.9%, respectively.
对一种以聚碳酸酯包被的铁珠为固相并配备磁处理装置的酶联免疫吸附测定法进行了评估,用于定量检测人血清样本中抗弓形虫抗体。在本研究确定的参数及其他基本条件下,该测定法具有高度可重复性:同日检测中阳性血清吸光度值的变异系数为2.42%,不同日检测中为3.75%。本测定系统与其他传统血清学检测之间存在显著相关性:与染色试验的相关系数为0.960,与乳胶凝集试验的相关系数为0.929。基于染色试验阳性和阴性血清样本吸光度值的频率分布进行的统计分析得出了区分阳性和可疑样本(0.357)以及可疑和阴性样本(0.266)的可行界限。在此诊断标准下,我们测定系统的结果与染色试验和乳胶凝集试验的结果非常吻合,一致性分别为94.9%和93.9%。