Overestimation of pediatric cardiac output by thermal indicator loss.

作者信息

Maruschak G F, Potter A M, Schauble J F, Rogers M C

出版信息

Circulation. 1982 Feb;65(2):380-3. doi: 10.1161/01.cir.65.2.380.

Abstract

The accurate measurement of pediatric cardiac output by thermodilution requires that the quantity of cold indicator introduced into the central circulation be known. This study defines an important source of error in the correction factor for the amount of heat gained by small volumes of cold injectate during passage through pediatric catheter systems. This error may result in significant overestimation of cardiac output (as much as 59%) when blood at body temperature is withdrawn into the injection lumen of the pediatric catheter before the injection.

摘要

通过热稀释法准确测量小儿心输出量需要知道引入中心循环的冷指示剂的量。本研究确定了一个重要的误差来源,即少量冷注射液在通过小儿导管系统时获得的热量的校正因子。当在注射前将体温的血液抽入小儿导管的注射腔内时,这种误差可能导致心输出量被显著高估(高达59%)。