Suppr超能文献

指示剂剂量、温度和心脏固有输出量会影响热稀释法测量心输出量的准确性和可重复性。

Indicator amount, temperature, and intrinsic cardiac output affect thermodilution cardiac output accuracy and reproducibility.

作者信息

Renner L E, Morton M J, Sakuma G Y

机构信息

School of Nursing, Department of Medicine, Oregon Health Sciences University, Portland 97201-3098.

出版信息

Crit Care Med. 1993 Apr;21(4):586-97. doi: 10.1097/00003246-199304000-00021.

Abstract

OBJECTIVE

To determine the accuracy and reproducibility of four thermodilution indicators (5-mL room temperature, 10-mL room temperature, 5-mL iced, and 10-mL iced injectates) at clinically relevant flow rates.

DESIGN

Quasi-experimental study.

SETTING

Animal research laboratory of a health sciences university.

SUBJECTS

Six virgin western-breed ewes.

INTERVENTIONS

Data were collected from six ewes that had ascending aorta electromagnetic flow probes and inferior vena cava occluders. Cardiac output was manipulated by inferior vena cava occlusion and isoproterenol infusion. Four thermodilution indicators were tested at high and low levels of cardiac output and compared with the electromagnetic flowmeter measurements of cardiac output.

MEASUREMENTS AND MAIN RESULTS

The indicator amounts were determined from both injectate volume and temperature difference between the injectate and blood. Using 5-mL room temperature injectate as a reference, 10-mL room contained 2 x, 5-mL iced 2.1 x, and 10-mL iced 4.1 x the indicator amount of 5-mL room temperature injectate. Approximately 210 simultaneous thermodilution and electromagnetic flow measurements were made for each injectate over a flow range of 1.5 to 15.7 L/min. For the entire cardiac output range, systematic error was not present. However, the r2 value (.92) for the 10-mL iced injectate group was greater (p < .05) than that value (.79) for the 5-mL iced injectate group, while r2 values were .79 for the 10-mL room temperature group and .49 for the 5-mL room temperature group. At flow rates of < 4.7 L/min, r2 was not different among injectates, but reduced indicator amounts progressively overestimated output (p < .05), reaching 21% for the 5-mL room temperature group. At flow rates of > 7.7 L/min, the r2 value (.81) for the 10-mL iced group was greater (p < .05) than that value (.45) for the 5-mL iced group, while r2 values were .24 for the 10-mL room temperature group and .08 for the 5-mL room temperature group. Systematic error was not present.

CONCLUSIONS

At low cardiac output levels, reduced indicator impairs accuracy but not reproducibility, a phenomenon that is perhaps related to indicator loss. At high cardiac output rates, reduced indicator impairs reproducibility. This phenomenon is probably related to low signal-to-noise ratio. Thermodilution indicator amounts should be tailored to the output range.

摘要

目的

确定四种热稀释指示剂(5毫升室温、10毫升室温、5毫升冰盐水和10毫升冰盐水注射液)在临床相关流速下的准确性和可重复性。

设计

准实验研究。

设置

一所健康科学大学的动物研究实验室。

对象

六只未生育的西方品种母羊。

干预措施

从六只安装了升主动脉电磁流量探头和下腔静脉封堵器的母羊收集数据。通过下腔静脉封堵和异丙肾上腺素输注来控制心输出量。在高、低心输出量水平下测试四种热稀释指示剂,并与心输出量的电磁流量计测量值进行比较。

测量和主要结果

根据注射液体积以及注射液与血液之间的温差确定指示剂含量。以5毫升室温注射液作为参考,10毫升室温注射液的指示剂含量是5毫升室温注射液的2倍,5毫升冰盐水注射液是2.1倍,10毫升冰盐水注射液是4.1倍。在1.5至15.7升/分钟的流速范围内,对每种注射液进行了约210次同步热稀释和电磁流量测量。在整个心输出量范围内,不存在系统误差。然而,10毫升冰盐水注射液组的r2值(0.92)大于(p < 0.05)5毫升冰盐水注射液组的值(0.79),而10毫升室温组的r2值为0.79,5毫升室温组为0.49。在流速< 4.7升/分钟时,各注射液之间的r2无差异,但指示剂含量减少会逐渐高估输出量(p < 0.05),5毫升室温组达到21%。在流速> 7.7升/分钟时,10毫升冰盐水组的r2值(0.81)大于(p < 0.05)5毫升冰盐水组的值(0.45),而10毫升室温组的r2值为0.24,5毫升室温组为0.08。不存在系统误差。

结论

在心输出量较低时,指示剂含量减少会损害准确性,但不影响可重复性,这种现象可能与指示剂损失有关。在心输出量较高时,指示剂含量减少会损害可重复性。这种现象可能与低信噪比有关。热稀释指示剂含量应根据输出范围进行调整。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验