Peterson P, Gottfried E L
Thromb Haemost. 1982 Apr 30;47(2):101-3.
The results of determinations of the prothrombin time (PT) and the activated partial thromboplastin time (aPTT) are frequently used to assess hemostatic function. Accurate results for these laboratory tests depend on many variables, one of which is the ratio of plasma to anticoagulant. We studied 12 patients and 4 normal subjects to determine the effects of sample volume on PT and aPTT. We conclude that underfilling may produce profound effects, particularly on the aPTT. In contrast, overfilling rarely affects the results. The greatest effects of sample volume were observed in specimens in which the true PT or aPTT was elevated. A normal PT or aPTT result on any specimen, regardless of sample volume, strongly suggests that the true value is normal.
凝血酶原时间(PT)和活化部分凝血活酶时间(aPTT)的测定结果常被用于评估止血功能。这些实验室检测的准确结果取决于许多变量,其中之一是血浆与抗凝剂的比例。我们研究了12名患者和4名正常受试者,以确定样本量对PT和aPTT的影响。我们得出结论,样本量不足可能会产生深远影响,尤其是对aPTT。相比之下,样本量过多很少影响结果。样本量的最大影响在真正的PT或aPTT升高的标本中观察到。任何标本上正常的PT或aPTT结果,无论样本量如何,都强烈表明真实值是正常的。