Alihanka J, Vaahtoranta K, Saarikivi I
Am J Physiol. 1981 May;240(5):R384-92. doi: 10.1152/ajpregu.1981.240.5.R384.
The principle of the static charge-sensitive-bed (SCSB) method is described. The method is simple and inexpensive. From the SCSB recording the ballistocardiogram (BCG) and respiratory movement can be simultaneously recorded by selective filtering of the original signal. The SCSB recording thus enables continuous long-term monitoring of the BCG, heart rate, respiratory rate, respiratory amplitude, and body movements. There are no electrodes or cables connected to the subject. The SCSB-BCG signal was studied by comparing the SCSB method with the conventional ultralow-frequency (ULF) acceleration BCG. The wave forms at rest and the amplitude responses to physical exercise were studied by both methods. Multiple recordings of the BCG, heart rate, respiratory movement, respiratory rate, and body movements after exercise and during sleep using the SCSB method are presented. The SCSB method opens new approaches to long-term studies of the regulation of myocardial performance, heart rate, and respiration. Applications of the SCSB method for clinical sleep studies, patient monitoring and cardiovascular screening examinations are discussed.
描述了静电荷敏感床(SCSB)方法的原理。该方法简单且成本低廉。通过对原始信号进行选择性滤波,利用SCSB记录可同时记录心冲击图(BCG)和呼吸运动。因此,SCSB记录能够对BCG、心率、呼吸频率、呼吸幅度和身体运动进行连续长期监测。无需将电极或电缆连接到受试者身上。通过将SCSB方法与传统的超低频(ULF)加速度BCG进行比较,对SCSB-BCG信号进行了研究。两种方法都研究了静息时的波形以及对体育锻炼的幅度响应。展示了使用SCSB方法对运动后和睡眠期间的BCG、心率、呼吸运动、呼吸频率和身体运动进行的多次记录。SCSB方法为长期研究心肌性能、心率和呼吸调节开辟了新途径。讨论了SCSB方法在临床睡眠研究、患者监测和心血管筛查检查中的应用。